| | Status: Tower installed, IVC pumping overnight. All mixture recovered to dump. 1K Pot needle valve closed. Cryostat vacuum jacket pumped out. Mixture cryocleaned once. |
| 0845 | Turbo = 40 nbar, pumping on IVC. IVC = 3.4 μbar with pump valved out (after 10 minutes). |
| 0850 | Open Trap 2 to G1: 340 mbar of contaminants after yesterday's 45 minute cryocleaning. Pump this out. |
| 0915 | Start cryocleaning mixture through Trap 2. V6=12.8% |
| 0915 | Quick electrical check. DIB4 pB Feedback coil/return reads open. This is a Side2 channel, so it is not critical. |
| 0925 | Cryocleaning flow stopped, increase V6 to 13.0% to reestablish circulation |
| 0925 | Leak detector is on IVC, background settles around 7.5 pbarL/s |
| 0930 | Do squirt-test leak check, no external IVC leaks found. |
| 1005 | Finish raising main bath |
| 1010 | Cryocleaning flow stopped, increase V6 to 13.1% to reestablish circulation |
| 1025 | Pump the mixture out of the trap and into the dump. Wait for G1 to stabilize. |
| 1038 | Close V12B+speedivalve, close V3, and close V6 completely. |
| 1040 | close V9, remove cold trap 2 from LN2 bath to warm up. |
| 1047 | Start cryocleaning mixture through trap 1. Open V6 to 13.1% |
| 1110 | Cryocleaning flow stopped, increase V6 to 13.3% to reestablish circulation |
| 1150 | Pump the mixture out of the trap and into the dump. Open V6. Wait for G1 to stabilize. End of cryocleaning through Trap 1, 60 minutes. |
| 1210 | Open Trap 2 (room temperature) to G1: 8.0 mbar of contaminants. Pump out with scroll pump. |
| 1225 | Verify 1K Pot needle valve closed. Pump out 1K Pot with scroll pump. |
| 1230 | Pump out fridge with scroll pump. There was a few mbar in the LHe cold trap, but no measurable amount in condenser or still. |
| 1235 | Open fridge lines up to (cold) Trap 2, to cryopump out whatever air is left in the fridge. |
| 1240 | Let mixture into fridge through Trap 2. 3He pump off, V9, V13b, V12b+speedi, V3, V1, condenser & still valves all open. |
| 1245 | Check IVC with leak detector: 11 pbarL/s. No significant leak from fridge to IVC. |
| 1250 | Pump out main bath with scroll pump. P2 falls below 1 mbar in 5 minutes. |
| 1300 | Pressurize main bath with He to 205 mbar above 1atm on G3. |
| note | He gas is left in the Vent volume, the auxiliary right volume, the auxiliary left (G3) volume, and the P2 volume (pump inlet). |
| 1305 | Let 5.4 mbar He exchange gas into IVC |
| 1310 | Begin LN2 transfer. |
| 1310 | MC(Cernox) = 295.7 K, Sorb = >249 K (max), IVC = 5.4 mbar, G1 (fridge) = 610 mbar, Level resistor = 0.988 kΩ. |
| 1320 | MC(Cernox) = 294.7 K, Sorb = >249 K (max), IVC = 4.6 mbar, G1 (fridge) = 610 mbar, Level resistor = 0.999 kΩ. |
| 1340 | MC(Cernox) = 263.4 K, Sorb = >249 K (max), IVC = 3.4 mbar, G1 (fridge) = 608 mbar, Level resistor = 1.033 kΩ. |
| 1400 | MC(Cernox) = 220.4 K, Sorb = >249 K (max), IVC = 2.5 mbar, G1 (fridge) = 607 mbar, Level resistor = 1.110 kΩ. |
| 1435 | MC(Cernox) = 156.3 K, Sorb = 240.5 K, IVC = 2.2 mbar, G1 (fridge) = 604 mbar, Level resistor = 1.265 kΩ. |
| 1435 | watching interchannel resistances on DIB4. pB is Side2, pC and pD are side 1. They were around 300 Ω at room temperature. Now pC-pD is around 20 kΩ and pB to the others is around 35 kΩ . |
| 1500 | MC(Cernox) = 131.0 K, Sorb = 174.1 K, IVC = 1.9 mbar, G1 (fridge) = 602 mbar, Level resistor = 1.308 kΩ. |
| 1520 | MC(Cernox) = 116.7 K, Sorb = 143.3 K, IVC = 1.9 mbar, G1 (fridge) = 601 mbar, Level resistor = 1.325 kΩ. |
| 1530 | MC(Cernox) = 111.1 K, Sorb = 133.8 K, IVC = 1.6 mbar, G1 (fridge) = 600 mbar, Level resistor = 1.330 kΩ. |
| | close to done with transfer. But for Run 64 we filled till the level resistor hit 1.347 kΩ, and it didn't completely hold overnight. |
| 1545 | MC(Cernox) = 104.1 K, Sorb = 125.1 K, IVC = 1.6 mbar, G1 (fridge) = 600 mbar, Level resistor = 1.335 kΩ. |
| 1545 | DIB4 pB - pC, pD around 400 MΩ, pC-pD around 1 MΩ |
| 1600 | MC(Cernox) = 98.1 K, Sorb = 120.7 K, IVC = 1.6 mbar, G1 (fridge) = 599 mbar, Level resistor = 1.337 kΩ. |
| 1615 | MC(Cernox) = 94.0 K, Sorb = 119.0 K, IVC = 1.6 mbar, G1 (fridge) = 598 mbar, Level resistor = 1.338 kΩ.
Stop LN2 transfer. |
| | after transfer, level resistor resistance dropped a bit, to 1.306 Ω. That's strange. Was it being cooled by gas flowing past it? Maybe not actually submerged? |
| 1625 | Perform electrical check around 95 K. A few issues found: |