| 1030 | Raise main bath. We had some trouble - the cryostat was tilted, so the bottom of the IVC kept getting caught on the lower rim in the bath. We adjusted one of the cable lengths to make it more level. It took a while to get the screw holes lined up but we eventually managed. |
| 1045 | Open fridge lines up to Trap 1, which is cold in the LN2 bath. This to let the little bit of air in the fridge get cryopumped out. |
| 1050 | Make sure 1K Pot needle valve is closed, then pump out 1K Pot with scroll pump. |
| 1110 | Connect leak detector to IVC and isolate turbo pump. LD signal is 1.3 pbarL/s. |
| 1115 | Let mixture into fridge through Trap 1. 3He pump off, V9, V13a, V12a+speedi, V3, V1, condenser & still valves all open. LD signal does not increase. |
| 1200 | Begin pumping out main bath. P2 falls below 100 mbar quickly. Gauge on MB line falls to 2“Hg (absolute) after a couple minutes. P2 falls below 1 mbar after 5 minutes. |
| 1210 | Pressurize MB with He gas to 140 mbar above 1atm on G3. LD signal drifts up slowly, 2.8 pbarL/s after a few minutes - good! Better than Run 63 and previous. |
| note | He gas is left in the Vent volume, the auxiliary right volume, the auxiliary left (G3) volume, and the P2 volume. |
| 1215 | LD has crept up to 5.1 pbarL/s |
| 1220 | Let 5.4 mbar He exchange gas into IVC |
| 1230 | Begin transferring LN2 to main bath |
| 1230 | MC(Cernox) = 296.6 K, Sorb = >249 K (max), IVC = 5.4 mbar, G1 (fridge) = 644 mbar, Level resistor = 1.066 kΩ. |
| 1240 | MC(Cernox) = 293.2 K, Sorb = >249 K (max), IVC = 5.4 mbar, G1 (fridge) = 642 mbar, Level resistor = 1.092 kΩ. |
| 1250 | MC(Cernox) = 277.1 K, Sorb = >249 K (max), IVC = 4.0 mbar, G1 (fridge) = 641 mbar, Level resistor = 1.135 kΩ. |
| 1300 | MC(Cernox) = 258.7 K, Sorb = >249 K (max), IVC = 4.0 mbar, G1 (fridge) = 640 mbar, Level resistor = 1.169 kΩ. |
| 1330 | MC(Cernox) = 200.0 K, Sorb = >249 K (max), IVC = 2.5 mbar, G1 (fridge) = 636 mbar, Level resistor = 1.330 kΩ. Nearly full. |
| 1335 | MC(Cernox) = 188.5 K, Sorb = 235.6 K, IVC = 2.5 mbar, G1 (fridge) = 636 mbar, Level resistor = 1.347 kΩ. End transfer. |
| 1445 | MC(Cernox) = 115.0 K, Sorb = 156.0 K, IVC = 1.9 mbar, G1 (fridge) = 636 mbar. Fridge pressure has been flat since end of transfer. |
| 1600 | MC(Cernox) = 95.7 K, Sorb = 151.4 K, IVC = 1.6 mbar, G1 (fridge) = 637 mbar. |
| 1630 | MC(Cernox) = 91.8 K, Sorb = 150.8 K, IVC = 1.6 mbar, G1 (fridge) = 638 mbar. |
| 1700 | Perform 77K electrical check. Mostly things look OK. But cross-channel resistances, through the qBias lines, look strange and inconsistent: |