| | Setup: DIB 1/3/4 → 3.2 MHz filters 1/2/3 → cable 1/3/2 → gender changers → rerouting boards 1/3/4 → DCRC55(1)/33(2)/50(3) |
| | Overnight baked for 5 hours. |
| 0931 | Tower/MC/Coldplate = 83.5/57.7/366 mK, Still = 1.26 K / 0.20 mbar, G1/G2 = 265/258 mbar, 1K Pot = 1.62 K / 26 mbar. MB = 25.3% |
| 0935 | Let's start with a repeat of the last series taken yesterday to see if baking vs flashing makes a difference. |
| 0936 | Start series of data with Vbias(external) = -15.00 V, with 1000 beginning randoms, and a 40/30/15/40/40 ADC bin threshold trigger on DCRC1qI/qO/DCRC2qO/DCRC3qI/qO, with trigger baseline = 32 samples and trigger length = 10 samples (these are 2x the usual values). |
| 0958 | Stop run, ground and flash for 60 s |
| 1013 | Start LHe transfer (MB = 24.0%) |
| 1045 | Fill LN2 cold trap. |
| 1121 | End LHe transfer (MB = 89.7%) |
| 1131 | Start series of data with Vbias(external) = -20.00 V, with 1000 beginning randoms, and a 40/30/15/40/40 ADC bin threshold trigger on DCRC1qI/qO/DCRC2qO/DCRC3qI/qO, with trigger baseline = 32 samples and trigger length = 10 samples (these are 2x the usual values). |
| 1153 | Stop run, ground and flash for 60 s |
| 1217 | Start series of data with Vbias(external) = -25.01 V, with 1000 beginning randoms, and a 40/30/15/40/40 ADC bin threshold trigger on DCRC1qI/qO/DCRC2qO/DCRC3qI/qO, with trigger baseline = 32 samples and trigger length = 10 samples (these are 2x the usual values). |
| 1238 | Stop run, ground and flash for 60 s |
| 1303 | Start series of data with Vbias(external) = -30.01 V, with 1000 beginning randoms, and a 40/30/15/40/40 ADC bin threshold trigger on DCRC1qI/qO/DCRC2qO/DCRC3qI/qO, with trigger baseline = 32 samples and trigger length = 10 samples (these are 2x the usual values). |
| 1324 | Stop run, ground and flash for 60 s |
| 1347 | Start series of data with Vbias(external) = -40.01 V, with 1000 beginning randoms, and a 40/30/15/40/40 ADC bin threshold trigger on DCRC1qI/qO/DCRC2qO/DCRC3qI/qO, with trigger baseline = 32 samples and trigger length = 10 samples (these are 2x the usual values). Noise Triggering ⇒ No DIB1 peaks |
| 1409 | Stop run, ground and flash for 60 s |
| 1433 | Start series of data with Vbias(external) = -50.00 V, with 1000 beginning randoms, and a 40/30/15/40/40 ADC bin threshold trigger on DCRC1qI/qO/DCRC2qO/DCRC3qI/qO, with trigger baseline = 32 samples and trigger length = 10 samples (these are 2x the usual values). Noise triggering ⇒ No DIB4 peaks |
| 1455 | Stop run, ground and flash for 60 s |
| 1516 | Start series of data with Vbias(external) = -70.0 V, with 1000 beginning randoms, and a 40/30/15/40/40 ADC bin threshold trigger on DCRC1qI/qO/DCRC2qO/DCRC3qI/qO, with trigger baseline = 32 samples and trigger length = 10 samples (these are 2x the usual values). Noise Triggering ⇒ No DIB1 peaks |
| 1538 | Stop run, ground and flash for 60 s |
| 1600 | Start series of data with Vbias(external) = -90.0 V, with 1000 beginning randoms, and a 40/30/15/40/40 ADC bin threshold trigger on DCRC1qI/qO/DCRC2qO/DCRC3qI/qO, with trigger baseline = 32 samples and trigger length = 10 samples (these are 2x the usual values). |
| 1622 | Stop run, ground and flash for 60 s |
| 1644 | Start series of data with Vbias(external) = -110.0 V, with 1000 beginning randoms, and a 40/30/15/40/40 ADC bin threshold trigger on DCRC1qI/qO/DCRC2qO/DCRC3qI/qO, with trigger baseline = 32 samples and trigger length = 10 samples (these are 2x the usual values). |
| 1706 | Stop run, ground detector |
| 1707 | Begin 12 hour bake |
| 1708 | Fill LN2 cold trap full. |