| | Setup: DIB 1/3/4 → 3.2 MHz filters 1/2/3 → cable 1/3/2 → gender changers → rerouting boards 1/3/4 → DCRC55(1)/33(2)/50(3) |
| | Overnight baked for 5 hours. |
| 0906 | Tower/MC/Coldplate = 83.7/57.6/370 mK, Still = 1.27 K / 0.20 mbar, G1/G2 = 266/253 mbar, 1K Pot = 1.62 K / 26 mbar. MB = 62.8% |
| 0911 | Fill LN2 cold trap. |
| 0912 | The dumps from the last few tests last night never were copied over to midasraw. Running ps -u cdmssoft shows that lazylogger is not running. Let's just get a clean slate and restart MIDAS. |
| 0917 | Now lazylogger is copying those series to midasraw. As suspected though, those series were both triggering on noise still. |
| 0920 | Start series of data with Vbias(external) = 5.01 V, with 1000 beginning randoms, and a 40/30/15/40/40 ADC bin threshold trigger on DCRC1qI/qO/DCRC2qO/DCRC3qI/qO, with trigger baseline = 32 samples and trigger length = 10 samples (these are 2x the usual values). |
| 0943 | Stop run, ground and flash for 60 s |
| 1004 | Start series of data with Vbias(external) = 10.01 V, with 1000 beginning randoms, and a 40/30/15/40/40 ADC bin threshold trigger on DCRC1qI/qO/DCRC2qO/DCRC3qI/qO, with trigger baseline = 32 samples and trigger length = 10 samples (these are 2x the usual values). |
| 1025 | Stop run, ground and flash for 60 s |
| 1046 | Start series of data with Vbias(external) = 15.00 V, with 1000 beginning randoms, and a 40/30/15/40/40 ADC bin threshold trigger on DCRC1qI/qO/DCRC2qO/DCRC3qI/qO, with trigger baseline = 32 samples and trigger length = 10 samples (these are 2x the usual values). |
| 1107 | Stop run, ground and flash for 60 s |
| 1128 | Start series of data with Vbias(external) = 20.00 V, with 1000 beginning randoms, and a 40/30/15/40/40 ADC bin threshold trigger on DCRC1qI/qO/DCRC2qO/DCRC3qI/qO, with trigger baseline = 32 samples and trigger length = 10 samples (these are 2x the usual values). |
| 1150 | Stop run, ground and flash for 60 s |
| 1212 | Start series of data with Vbias(external) = 25.00 V, with 1000 beginning randoms, and a 40/30/15/40/40 ADC bin threshold trigger on DCRC1qI/qO/DCRC2qO/DCRC3qI/qO, with trigger baseline = 32 samples and trigger length = 10 samples (these are 2x the usual values). |
| 1245 | Stop run, ground and flash for 60 s |
| 1309 | Experiencing some PAN network errors which are messing with MIDAS. |
| 1319 | Start series of data with Vbias(external) = 40.00 V, with 1000 beginning randoms, and a 40/30/15/40/40 ADC bin threshold trigger on DCRC1qI/qO/DCRC2qO/DCRC3qI/qO, with trigger baseline = 32 samples and trigger length = 10 samples (these are 2x the usual values). |
| 1343 | Stop run, ground and flash for 60 s |
| 1403 | We took good 50,70,90 V data the other day. |
| 1404 | Start series of data with Vbias(external) = 110.1 V, with 1000 beginning randoms, and a 40/30/15/40/40 ADC bin threshold trigger on DCRC1qI/qO/DCRC2qO/DCRC3qI/qO, with trigger baseline = 32 samples and trigger length = 10 samples (these are 2x the usual values). |
| 1428 | Stop run. Measure Leakage voltage of 0.00 mV using Fluke 8600A. |
| 1430 | Begin 10 minute mini-bake. |
| 1518 | Tower = 83.9 mK |
| 1519 | Start series of data with Vbias(external) = -5.009 V, with 1000 beginning randoms, and a 40/30/15/40/40 ADC bin threshold trigger on DCRC1qI/qO/DCRC2qO/DCRC3qI/qO, with trigger baseline = 32 samples and trigger length = 10 samples (these are 2x the usual values). |
| 1540 | Stop run, ground and flash for 60 s |
| 1600 | Start series of data with Vbias(external) = -10.00 V, with 1000 beginning randoms, and a 40/30/15/40/40 ADC bin threshold trigger on DCRC1qI/qO/DCRC2qO/DCRC3qI/qO, with trigger baseline = 32 samples and trigger length = 10 samples (these are 2x the usual values). |
| 1649 | Stop run, ground and flash for 60 s |
| 1709 | Start series of data with Vbias(external) = -15.00 V, with 1000 beginning randoms, and a 40/30/15/40/40 ADC bin threshold trigger on DCRC1qI/qO/DCRC2qO/DCRC3qI/qO, with trigger baseline = 32 samples and trigger length = 10 samples (these are 2x the usual values). |
| 1729 | Stop run, ground detector. |
| 1730 | Start 12 hour bake. |
| 1731 | LN2 tank runs dry. Will need to fill this in the morning! |