start › cdms › k100 › run_diary › 20171024

2017/10/24 (Tuesday)

Previous Day

Setup: DIB 1/3/4 → 3.2 MHz filters 1/2/3 → cable 1/3/2 → gender changers → rerouting boards 1/3/4 → DCRC55(1)/33(2)/50(3)
Overnight baked for 5 hours.
0906 Tower/MC/Coldplate = 83.7/57.6/370 mK, Still = 1.27 K / 0.20 mbar, G1/G2 = 266/253 mbar, 1K Pot = 1.62 K / 26 mbar. MB = 62.8%
0911 Fill LN2 cold trap.
0912 The dumps from the last few tests last night never were copied over to midasraw. Running ps -u cdmssoft shows that lazylogger is not running. Let's just get a clean slate and restart MIDAS.
0917 Now lazylogger is copying those series to midasraw. As suspected though, those series were both triggering on noise still.
0920 Start series of data with Vbias(external) = 5.01 V, with 1000 beginning randoms, and a 40/30/15/40/40 ADC bin threshold trigger on DCRC1qI/qO/DCRC2qO/DCRC3qI/qO, with trigger baseline = 32 samples and trigger length = 10 samples (these are 2x the usual values).
0943 Stop run, ground and flash for 60 s
1004 Start series of data with Vbias(external) = 10.01 V, with 1000 beginning randoms, and a 40/30/15/40/40 ADC bin threshold trigger on DCRC1qI/qO/DCRC2qO/DCRC3qI/qO, with trigger baseline = 32 samples and trigger length = 10 samples (these are 2x the usual values).
1025 Stop run, ground and flash for 60 s
1046 Start series of data with Vbias(external) = 15.00 V, with 1000 beginning randoms, and a 40/30/15/40/40 ADC bin threshold trigger on DCRC1qI/qO/DCRC2qO/DCRC3qI/qO, with trigger baseline = 32 samples and trigger length = 10 samples (these are 2x the usual values).
1107 Stop run, ground and flash for 60 s
1128 Start series of data with Vbias(external) = 20.00 V, with 1000 beginning randoms, and a 40/30/15/40/40 ADC bin threshold trigger on DCRC1qI/qO/DCRC2qO/DCRC3qI/qO, with trigger baseline = 32 samples and trigger length = 10 samples (these are 2x the usual values).
1150 Stop run, ground and flash for 60 s
1212 Start series of data with Vbias(external) = 25.00 V, with 1000 beginning randoms, and a 40/30/15/40/40 ADC bin threshold trigger on DCRC1qI/qO/DCRC2qO/DCRC3qI/qO, with trigger baseline = 32 samples and trigger length = 10 samples (these are 2x the usual values).
1245 Stop run, ground and flash for 60 s
1309 Experiencing some PAN network errors which are messing with MIDAS.
1319 Start series of data with Vbias(external) = 40.00 V, with 1000 beginning randoms, and a 40/30/15/40/40 ADC bin threshold trigger on DCRC1qI/qO/DCRC2qO/DCRC3qI/qO, with trigger baseline = 32 samples and trigger length = 10 samples (these are 2x the usual values).
1343 Stop run, ground and flash for 60 s
1403 We took good 50,70,90 V data the other day.
1404 Start series of data with Vbias(external) = 110.1 V, with 1000 beginning randoms, and a 40/30/15/40/40 ADC bin threshold trigger on DCRC1qI/qO/DCRC2qO/DCRC3qI/qO, with trigger baseline = 32 samples and trigger length = 10 samples (these are 2x the usual values).
1428 Stop run. Measure Leakage voltage of 0.00 mV using Fluke 8600A.
1430 Begin 10 minute mini-bake.
1518 Tower = 83.9 mK
1519 Start series of data with Vbias(external) = -5.009 V, with 1000 beginning randoms, and a 40/30/15/40/40 ADC bin threshold trigger on DCRC1qI/qO/DCRC2qO/DCRC3qI/qO, with trigger baseline = 32 samples and trigger length = 10 samples (these are 2x the usual values).
1540 Stop run, ground and flash for 60 s
1600 Start series of data with Vbias(external) = -10.00 V, with 1000 beginning randoms, and a 40/30/15/40/40 ADC bin threshold trigger on DCRC1qI/qO/DCRC2qO/DCRC3qI/qO, with trigger baseline = 32 samples and trigger length = 10 samples (these are 2x the usual values).
1649 Stop run, ground and flash for 60 s
1709 Start series of data with Vbias(external) = -15.00 V, with 1000 beginning randoms, and a 40/30/15/40/40 ADC bin threshold trigger on DCRC1qI/qO/DCRC2qO/DCRC3qI/qO, with trigger baseline = 32 samples and trigger length = 10 samples (these are 2x the usual values).
1729 Stop run, ground detector.
1730 Start 12 hour bake.
1731 LN2 tank runs dry. Will need to fill this in the morning!

Next Day