| | Status: New oil in 3He pump. Pumped mixture out of cryopump and into dump overnight. Still a little mixture in Trap 2. |
| 1725 | P1 (cryopump) = 0.180 mbar and falling. G2 (dump) = 738.5 mbar and rising. |
| 0850 | P1 (cryopump) = 0.034 mbar. Steady fall overnight, pretty flat now. G2 (dump) = 732.0 mbar. This has decreased overnight, possibly just due to change in atmospheric pressure. It's lower than the ~780 mbar level we had before the oil change it seems. Did we lose a little mixture? Should do a careful comparison. |
| 0850 | 3He Pump oil level = 40%. Bouncy meniscus, some bubbles. |
| 0855 | Close black valve on cryopump. P1 falls a little, to 0.020 mbar. |
| 0900 | Close Vent speedivalve. P1 falls slightly, to 0.018 mbar. We have now pumped out the cryopump. |
| 0900 | Open V12b+speedivalve to view Trap 2 pressure on G1. G1 actually falls from 5 to 2 mbar. (More air had leaked into IGH than there was mix still in the trap.) |
| 0905 | Close V7, open V2. Now pumping out Trap 2 to dump through V12b+speedi, V2, V8, V6, V9. |
| 0905 | G1/P1 (Trap 2) → 1.8/0.300 mbar. G2 (dump) increases, 732→735 mbar. |
| 0940 | G1/P1 (Trap 2) = 1.9/0.260 mbar. Pretty flat. G2 (dump) = 734.3 mbar. (G2 falling at same rate as before, possibly due to the weather.) |
| 1050 | G1/P1 (Trap 2) = 1.8/0.200 mbar. Pretty flat. G2 (dump) = 733.3 mbar. This is certainly long enough. Stop pumping out Trap 2. |
| 1100 | Prepare to cryoclean mixture. Put Cold Trap 2 in bath and fill bath with LN2. |
| 1110 | Begin cryocleaning mixture through Trap 2. V6 open to 12.8%. |
| 1115 | Flow stopped, increase V6 to 13.1% to re-establish flow. |
| 1125 | Flow stopped, increase V6 to 13.4% to re-establish flow. This was too much - close V6, then restart. Open V6 to 13.1% to establish flow. |
| 1135 | Flow stopped, increase V6 to 13.3% to re-establish flow. |
| 1155 | Stop cryocleaning after 45 minutes. |
| 1225 | Trap 2 is pumped out. G2(dump) = 734.0 mbar, pretty flat. Remove Trap 2 from bath to let warm up. Put Trap 1 into bath for a second cleaning. |
| 1230 | Begin cryocleaning mixture through Trap 1. V6 open to 13.3%. |
| 1235 | Flow stopped (or very slow), increase V6 to 13.4% to re-establish flow. |
| 1315 | Oops - looks like I had V9 open during the cleaning. So wasn't really flowing through the trap. |
| 1320 | Begin cryocleaning mixture through Trap 1. V6 open to 13.1%. This time V9 is closed as it should be. |
| 1330 | Flow stopped, increase V6 to 13.2% to re-establish flow. |
| 1400 | Flow stopped, increase V6 to 13.3% to re-establish flow. |
| 1410 | Stop cryocleaning after 50 minutes. |
| 1440 | Trap 1 is pumped out. G2(dump) = 730.0 mbar, rising slowly. Remove Trap 1 from bath to let warm up. |
| 1440 | Open (warm) Trap 2 to G1. G1 = 331 mbar. That's a lot of contamination! Pump out with scroll pump. |
| 1455 | Replace Trap 2 in LN2 bath. |
| 1500 | Begin cryocleaning mixture through Trap 2. V6 open to 13.0%. |
| 1500 | Flow stopped, increase V6 to 13.1% to re-establish flow. |
| 1505 | Flow stopped, increase V6 to 13.2% to re-establish flow. |
| 1530 | Flow stopped, increase V6 to 13.3% to re-establish flow. |
| 1550 | Stop cryocleaning after 50 minutes. |
| 1615 | Trap 2 is pumped out. G2(dump) = 733.7 mbar, rising slowly. Remove Trap 2 from bath to let warm up. |
| 1620 | Open (warm) Trap 1 to G1. G1 = 5.0 mbar. That's a lot less than last time. Pump out with scroll pump. |
| 1640 | Put Trap 1 back in LN2 bath. |
| 1705 | G2 (dump, with 3He pumping for a while, V6 closed) = 734.5 mbar. Close V9 and turn off pump. |
| 1620 | Trap 2 is nearly room temperature and open to G1. G1 = 5.2 mbar. Pump out with scroll pump. |