Plug DCRC1 into mini-bob and plug in power
connect via telnet
rd 0: this is updating, different value every time. not stuck
rd 14: this reads 0. rd 2 and rd 3 are 0.
wr 2 00F0, wr 3 1728. Then rd 14, non-zero value. rd 14 repeatedly - it's value keeps changing.
On ODB enable only DCRC1, set as master. Then start programs. rd 0 and rd 14 are still updating. wr 2 and wr 3 have the same values as I entered.
start run with MIDAS, then check rd 0. Updates for 20 seconds, then gets stuck.
also rd 14 is stuck. rd 2 is updating
stop run, rd 2 is no longer updating. rd 0 and rd 14 still stuck.
rd 2 = 00E5, rd 3 = 4CA8. Try changing these - the changes register, but rd 14 and rd 0 don't change.
stop programs. DCRC is still stuck.
power cycle DCRC.
repeat 1-7. DCRC takes longer to get stuck, somewhere between 3 and 4 minutes.