UMN CDMS Wiki archive
start › cdms › k100 › run_diary › 20170908

2017/09/08 (Friday)

Previous Day

Status Both LN2 traps warm and pumped out. Mixture in dump (LHe cold trap isolated at around 20 K). 1K Pot needle valve open. Cryostat vacuum jacket pumped out.
0955 Perform wirebond repair on S1502 qD (DIB3 qO). Used power 480/480, time 40/30, length/drop 50/26. Didn't stick to the detector pad at first, had to open/feed more wire out, then it stuck after “bond off excess”. Resistance checks look good.
1000 Examined G132 Side1 pF. This is the channel that seemed to have no connection issues but behaved as if open/high R when cold. Looked at it under a microscope and took some pictures. No obvious problems with the traces or wirebond leading from the wirebond pads to the center phonon channel.
1332 Measuring detector-electrode gap. Used micrometer and flat bar to measure distance from delector/electrode face to housing rims. Measurements indicated the outer rim is where adjascent housings meet.

Housing Measurement Length (inches ±0.001)
electrode bar to electrode 0.3835
electrode bar to inner rim 0.4182
electrode bar to outer rim 0.4560
crystal bar to crystal 0.4940
crystal bar to inner rim 0.4290
crystal bar to outer rim 0.3842
Using the outer rim as reference, the electrode is 0.4560“-0.3835”=0.0725“ above and the crystal face is 0.4940”-0.3842“=0.1098” above the outer rim (when stack is upright). This gives us a gap of 0.037“±0.004” between detector and electrode. This is just under 1mm.
1400 Bill V was able to drill a 6 mil hole in the center of one of our extra 3/8“ lead disks. So we are switching out the 8 mil collimator on qB's source with a 6 mil collimator.
1435 Assemble stack. Crown, then electrode housing (electrode facing up), then detector housing (bare face downward), then lid, then thermometer over one of the screw holes, then the 6 threaded rods made for this run.
1500 Proceeded with electrical check at the DIBs but some cross-channel resistances were unusually low. So we took the lid off an measured from pin to channel (second lead directly on detector)
Detector Electrical Checkout (open housing) Date: 2017 Sep 8
Detector: S1502 Who: MF NM
Pin qA qB qC qD qE
DIB1 qI (qA) 2 16 200 200 14
DIB4 qI (qB) 14 2 200 200 30
DIB1 qO (qC) 200 200 3 200 200
DIB3 qO (qD) 200 200 200 4 200
DIB4 qO (qE) 11 20 200 200 5
all measurements in Ω. (for >100 approximate values shown)

Examining table above suggests 1 short between qA and qB and another between qA and qE. Looking closely at detector, one bond seemed to be touching the detector surface - gently pushed it up with scalpel. For the other short it looked like a tiny piece of grit was between the T-shaped bond pad and the adjacent channel - removed this with rolled up piece of wipe dampened with methanol.

Detector Electrical Checkout (open housing) Date: 2017 Sep 8
Detector: S1502 Who: MF NM
Pin qA qB qC qD qE
DIB1 qI (qA) 2 200 200 300 150
DIB4 qI (qB) 200 2 200 300 300
DIB1 qO (qC) 300 200 3 300 300
DIB3 qO (qD) 300 300 300 4 300
DIB4 qO (qE) 100 300 300 300 5
all measurements in Ω. (for >100 approximate values shown)

This seems to have fixed the shorts!

1525 Reassemble stack. Unfortunately the threads in the crown seem to be getting worse…
Detector Electrical Checkout Date: 2017 Sep 8
Detector: S1502 Who: MF NM
+lead -lead Measurement Value
DIB1 qO DIB1 qI qC to qA 300 Ω
DIB1 qI DIB3 qO qA to qD 500 Ω
DIB4 qO DIB4 qI qE to qB 700 Ω
DIB4 qO DIB1 qO qE to qC 640 Ω
DIB3 qO DIB4 qI qD to qB 480 Ω
DIB1 qO chassis <20 nS
DIB1 qI chassis 10 nS
DIB3 qO chassis 8.5 MΩ
DIB4 qO chassis 8.5 MΩ
DIB4 qI chassis 8.5 MΩ
note: qI is the last pin on the DIB
1630 Start building recovery gas warming tube - 50' of 1/2” OD copper “refrigeration tubing” and two brass 25mm KF blanks. Drilled 3/8“ hole and 1/2” flat depression into blank then solder onto tube.

Next Day