2017/09/19 (Tuesday)

Previous Day

Status: Tower installed and has passed electrical checks. IVC passed leak check. Pumped out IVC with turbo pump since Friday.

Decision on He-3 pump:

  • We will try using the pump normally for this run. After the run we will change the oil. This requires pulling mixture out of the pump with the cryopump. We will use 3L of LVO 210 pump oil (this oil was used on 2014 November 20).
  • The second pump is leaky according to Hassan's exit email from June 2015. We should contact Oerlikon Leybold for maintenance/repair/replacement.
1115 Turbo = 22 nbar, IVC = 0.74 μbar.
1120 Check the cold traps: each has about 1 mbar. Pump them out with scroll pump. Then fill LN2 cold trap bath.
1130 Begin cleaning mixture through Cold Trap 2 with V6 open to 12.6%.
1135 Circulation stopped, increase V6 to 12.8% to restart circulation
Notice that pump is louder / whinier now than earlier today, with a larger inlet pressure.
1210 Circulation stopped, increase V6 to 13.0% to restart circulation
1300 Stop cryocleaning after 90 minutes, Trap 2
1350 Warm up Trap 2 and pump out: 62.5 mbar of contaminants.
1405 Begin cleaning mixture through Cold Trap 2 with V6 open to 12.3%. Flow stops, increase V6 to 12.8% to start flow again.
1410 Circulation stopped, increase V6 to 13.0% to restart circulation
1540 Stop cryocleaning after 95 minutes, Trap 2
1555 Begin cleaning mixture through Cold Trap 1 with V6 open to 13.1%.
1610 Circulation stopped, increase V6 to 13.2% to restart circulation
1705 Stop cryocleaning after 70 minutes, Trap 1