2017/09/05 (Tuesday)
| Status | Both LN2 traps warm and pumped out. Mixture in dump (LHe cold trap isolated at around 20 K). 1K Pot needle valve open. Cryostat vacuum jacket pumped out. |
|---|---|
| 1115 | Post-run 63 tower electrical checkout: |
| Detector | G132 | DIB | 1 | ||
|---|---|---|---|---|---|
| side coax | ST-3-3 | date/who: | 2017-09-05 | GR MF | |
| face | 1 | ||||
| +lead | - lead | signal | typical | prev. | measured |
| 1 | 2 | QET A | 100 to 200 Ω | 174 Ω | 174 Ω |
| 2 | 4 | QETA-B | 100 to 200 Ω | 209 | 206 Ω |
| 3 | 4 | QET B | 100 to 200 Ω | 184 Ω | 183 Ω |
| 4 | 6 | QETB-C | 100 to 200 Ω | 184 | 183 Ω |
| 5 | 6 | QET C | 100 to 200 Ω | 157 Ω | 157 Ω |
| 6 | 8 | QETC-D | 100 to 200 Ω | 172 | 170 Ω |
| 7 | 8 | QET D | 100 to 200 Ω | 97.5 Ω | 98 Ω |
| 2 | 8 | QETA-D | 100 to 200 Ω | 179 | 177 Ω |
| 9 | chassis | LED1 Rforward | 8 to 9 MΩ | 730 kΩ | 680 kΩ |
| 9 | chassis | LED1 Vforward | 1.0 to 1.1 V | 1.18 V | 1.118 V |
| chassis | 9 | LED1 Rreverse | <1 nS | 1.5 nS | 2 nS |
| 10 | chassis | LED2 Rforward | 8 to 9 MΩ | 740 kΩ | 740 kΩ |
| 10 | chassis | LED2 Vforward | 1.0 to 1.1 V | 1.11 V | 1.15 V |
| chassis | 10 | LED2 Rreverse | <1 nS | 1.8 nS | 2 nS |
| 1 | 11 | Qo Rbias | 24 to 26 nS | 25 nS | 25 nS |
| 12 | 11 | Qo C coupling | <1 nS | <1 nS | <1 nS |
| 12 | 13 | Qo Rfeedback | 24 to 26 nS | 25 nS | 25 nS |
| 1 | 14 | Qi Rbias | 24 to 26 nS | 26 nS | 25 nS |
| 15 | 14 | Qi C coupling | <1 nS | <1 nS | <1 nS |
| 15 | 16 | Qi Rfeedback | 25 nS | 25 nS | |
| All pins → Ground | <1.0nS | short | 1-8 ~150 Ω, 9-10 ~700 kΩ | ||
| Detector | G132 | DIB | 3 | ||
|---|---|---|---|---|---|
| side coax | ST-3-9 | date/who: | 2017-09-05 | GR MF | |
| face | 5 | ||||
| +lead | - lead | signal | typical | prev. | measured |
| 1 | 2 | QET A | 100 to 200 Ω | 114 Ω | 114 Ω |
| 2 | 4 | QETA-B | 100 to 200 Ω | 161 | 159 Ω |
| 3 | 4 | QET B | 100 to 200 Ω | 129 Ω | 129 Ω |
| 4 | 6 | QETB-C | 100 to 200 Ω | 186* Ω | open |
| 5 | 6 | QET C | 100 to 200 Ω | 153* Ω | open |
| 4 | 5 | QETB-C | 100 to 200 Ω | 169 | 166 Ω |
| 6 | 8 | QETC-D | 100 to 200 Ω | 260 | open |
| 7 | 8 | QET D | 100 to 200 Ω | 132 Ω | 133 Ω |
| 2 | 8 | QETA-D | 100 to 200 Ω | 167 | 165 Ω |
| 9 | chassis | LED1 Rforward | 8 to 9 MΩ | 336 | 980 kΩ |
| 9 | chassis | LED1 Vforward | 1.0 to 1.1 V | 0.65! V | ol |
| chassis | 9 | LED1 Rreverse | <1 nS | 15 MΩ | 16 MΩ |
| 10 | chassis | LED2 Rforward | 8 to 9 MΩ | 793 kΩ | 790 kΩ |
| 10 | chassis | LED2 Vforward | 1.0 to 1.1 V | 1.11 V | 1.116 V |
| chassis | 10 | LED2 Rreverse | <1 nS | <1 ns | <1 nS |
| 1 | 11 | Qo Rbias | 24 to 26 nS | 25 nS | 25 nS |
| 12 | 11 | Qo C coupling | <1 nS | <1 nS | <1 nS |
| 12 | 13 | Qo Rfeedback | 24 to 26 nS | 25 nS | 25 nS |
| 1 | 14 | Qi Rbias | 24 to 26 nS | 25 nS | 25 nS |
| 15 | 14 | Qi C coupling | <1 nS | <1 nS | <1 nS |
| 15 | 16 | Qi Rfeedback | 24 to 26 nS | 25 nS | 25 nS |
| All pins → Ground | <1.0nS | short, ~MΩ 1-8. Later worrse | 1-8 ~100Ω except 6 open. Others open | ||
| * = this was open at first. Pushed on the side coax at dib to fix it ! = here's the 1650 nm LED after we found the shorts on this DIB we found them on DIB 1 as well, originally that one was ok |
|||||
| Detector | G132 | DIB | 4 | ||
|---|---|---|---|---|---|
| side coax | ST-3-10 | date/who: | 2017-09-05 | GR MF | |
| face | 3 | ||||
| +lead | - lead | signal | typical | prev. | measured |
| 1 | 2 | QET A | 100 to 200 Ω | 143 Ω | 143 Ω |
| 2 | 4 | QETA-B | 100 to 200 Ω | 188 Ω | 189 Ω |
| 3 | 4 | QET B | 100 to 200 Ω | 185 Ω | 186 Ω |
| 4 | 6 | QETB-C | 100 to 200 Ω | 200 Ω | 204 Ω |
| 5 | 6 | QET C | 100 to 200 Ω | 181 Ω | 182 Ω |
| 6 | 8 | QETC-D | 100 to 200 Ω | 190 Ω | 191 Ω |
| 7 | 8 | QET D | 100 to 200 Ω | 134 Ω | 132 Ω |
| 2 | 8 | QETA-D | 100 to 200 Ω | 174 Ω | 175 Ω |
| 9 | chassis | LED1 Rforward | 8 to 9 MΩ | 794 kΩ | 790 kΩ |
| 9 | chassis | LED1 Vforward | 1.0 to 1.1 V | 1.11 V | 1.12 V |
| chassis | 9 | LED1 Rreverse | <1 nS | <1 nS | 1 nS |
| 10 | chassis | LED2 Rforward | 8 to 9 MΩ | 786 kΩ | 790 kΩ |
| 10 | chassis | LED2 Vforward | 1.0 to 1.1 V | 1.10 V | 1.2 V |
| chassis | 10 | LED2 Rreverse | <1 nS | <1 nS | 1 nS |
| 1 | 11 | Qo Rbias | 24 to 26 nS | n/c | <1 nS |
| 12 | 11 | Qo C coupling | <1 nS | <1 nS | <1 nS |
| 12 | 13 | Qo Rfeedback | 24 to 26 nS | 26 nS | 26 nS |
| 1 | 14 | Qi Rbias | 24 to 26 nS | n/c | n/c |
| 15 | 14 | Qi C coupling | <1 nS | <1 nS | <1 nS |
| 15 | 16 | Qi Rfeedback | 24 to 26 nS | 26 nS | 26 nS |
| All pins → Ground | <1.0nS | 1-8 ~100-400 Ω | |||
previous:
| Pins to chassis [Ω] | ||||||||
|---|---|---|---|---|---|---|---|---|
| 1 | 2 | 3 | 4 | 5 | 6 | 7 | 8 | |
| DIB1 | 132 | 127 | 126 | 124 | 122 | 121 | 95 | 96 |
| DIB3 | 53 | 61 | 107 | 104 | 106 | 115 | 114 | 111 |
| DIB4 | 114 | 113 | 123 | 124 | 123 | 127 | 115 | 112 |
now:
| Pins to chassis [Ω] | ||||||||
|---|---|---|---|---|---|---|---|---|
| 1 | 2 | 3 | 4 | 5 | 6 | 7 | 8 | |
| DIB1 | 132 | 127 | 126 | 124 | 122 | 121 | 95 | 96 |
| DIB3 | 53 | 61 | 107 | 104 | 106 | 115 | 114 | 111 |
| DIB4 | 114 | 113 | 123 | 124 | 123 | 127 | 115 | 112 |
| 1155 | On DIB3pC (5-6), which is open, we were able to get it to connect by pushing in the side coax at the detector dib. ~120 Ω. Becomes open again when you stop pushing on it. |
|---|---|
| 1200 | Continue tower disassembly |
| 1215 | While removing side coax ST-3-9, hit wirebond area with thumb. Probably broke some bonds… yes, seems like all resistor connections are broken, maybe one of the capacitor connections. Will need to repair with wirebonder. Althrough we only need two to run charge on the iZIP, so could wait on this. |
| 1245 | Put G132 in purge. NOTE: we forgot to put the tower in purge after deinstallation, so it has been in the anteroom for a week. |