2017/07/19 (Wednesday)
| Setup: DIB 1/3/4 → 3.2 MHz filters 1/2/3 → cable 1/2/3 → gender changer 1/3/4 → DCRC55(1)/33(2)/X(4) | |
| 0830 | Tower = 25.9 mK, Still = 0.060 mbar, G1/G2 = 182/166 mbar, 1K Pot = 1.716 K / 40 mbar. MB = 68.9% (predicted level: 72.1%) |
| 0830 | Lab = 23.4 C, Humidity = 41% |
| 0835 | Fill LN2 cold trap bath. Dewar empty, not sure whether it filled completely. |
| Overnight: took charge-bias-scan data. Last flash occurred at 00:26 am. No indication of elevated tower temperatures after biasing. Actually it looks like a DAQ crash occurred. The 00:26 flash would have been after the 5th series, at -1/+1 V. /byseries contains 5 series starting yesterday at 17:41, with the correct amount of dumps. So it looks like the crash occurred during the cooldown period between 00:26 and 00:51. MIDAS web page is down. |
cdms2:online cdmssoft$ps -u cdmssoft
PID TTY TIME CMD 14568 pts/0 00:00:00 bash 14773 ? 00:00:09 mserver 14775 ? 01:04:18 mlogger 14779 ? 00:03:27 lazylogger 14860 ? 00:00:51 dcrc_driver.exe 14863 ? 06:07:52 towerfe3.exe 14865 ? 09:59:38 triggerfe2.exe 32369 ? 00:01:24 mserver 63692 pts/0 00:00:00 ps
Report entered on crash report page
| As seen above the http process is not running. We'll try to kill then restart MIDAS. | |
| 0850 | Execute the kill script. This stopped most of the programs but not dcrc_driver, towerfe3, or triggerfe2. Try executing the start script. |
| 0850 | This seems to have worked. Sequence has restarted itself. It appears to be taking data now. A new series has begun, 07170719_0849. DCRC Control shows bias of -1.5/1.5 V. Difficult to say whether bias was set just now or hours ago, so not certain whether we're in a good neutralization state. But probably it was set just now, and the sequence can continue normally. |
| 0855 | Looking at BatViewer it appears that triggers are happening too frequently, very few pulses are in the right place. |
| 0856 | Increase DCRC1 q[0] threshold from 45 to 100 |
| 0905 | Check BatViewer again, pulse rate looks reasonable. Checking previous series, the first 4 look ok - but the 5th one at -1/+1 looks like mostly random triggers. Possibly the noise state of DCRC1 q[0] changed. |
| 1100 | Set Sequencer to terminate after -2/2V series(07170719_1015). |
| 1129 | Flash detector for 30s. |
| 1246 | mhttp seems to have crashed again. Repeating the above procedure restores it. |
| 1250 | Realize that terminating the above sequence early left the detector biased. Turn off bias in odb and flash again. |
| 1335 | Tower/MC/Coldplate = 26.6/30.3/184 mK, Still = 1.23 K / 0.065 mbar, G1/G2 = 188/173 mbar, 1K Pot = 1.71 K / 38 mbar. |
| 1352 | Based on preliminary Charge Collection Efficiency results and phonon baseline shifts with voltage bias, decide on Side1/2=-/+ 2V bias for this detector. |
| 1502 | Fill LN2 cold trap full. |
|---|---|
| 1525 | Complete DCRC1PA IbIs at 2/-2 V bias. |
| 1700 | Complete DCRC1PC IbIs at 2/-2 V bias. |
| 1711 | Fill LN2 cold trap full. |
| 1852 | Take short test series triggering on testSignal, with testSignal connected to DCRC1PA with 100Hz, DDSADCScaling=4 |
| 1859 | Begin test frequency scan sequence, “testSigScan_log.msl”. Not writing as cdms series, series numbers 7855-7885. |