| 1247 | Tower/MC/Coldplate = 26.7/29.5/187 mK, Still = 1.22 K / 0.065 mbar, G1/G2 = 183/171 mbar, 1K Pot = 1.74 K / 50 mbar. MB = 22.9% (Predicted level: 27.9%) |
| 1305 | Start LHe transfer (MB = 22.5%) |
| 1359 | End LHe transfer (MB = 92.2%) |
| 1415 | Release pressure from MB. |
| 1625 | Fill LN2 cold trap full. |
| 1644 | Modified charge_bias_scan.msl in a text editor for negative biases. Instead of using the Sequencer “load” button, I simply refreshed the page which seemed to update the sequence script on the web page. However, when I tried to start the sequence, it behaved strangely, skipping parts of the code. Aborted the sequence with LEDs all turned on. Turn these off manually. |
| 1645 | After loading the sequence properly using the button, it seems to work fine now. |
| 1721 | Waiting for the fridge to cool from that aborted series but it seems to be plateauing at 32mK. Realize that DCRC4 15VPower was left on when it is usually off. Turn it off in the ODB and temp quickly falls to 27.5mK and continues to cool. |
| 1731 | Start sequence 12 60-minute data series of charge-bias-scan data with Frontend Gain = (15/2)/(4/4)/(5/5/5/5), OG = 4/2/4, QET biases = 2017/07/17, with 1000 beginning randoms, 30 sec flash with 25 min wait, and a 45ADC bin threshold trigger on DCRC1 q[0]. Charge biases are DCRC1/2 = -/+ = Side1/2 = +/-. Side 1 biases in order are 0, 0.25, 0.5, 0.75, 1, 1.5, 2, 2.5, 3, 3.5, 4, 5 |
| | Note: This data was started with a 10 minute delay from start time. |