start › cdms › k100 › run_diary › 20170614

2017/06/14 (Wednesday)

Previous Day

Status Began warmup yesterday. Mixture has been pumped out of LN2 traps, which are currently warm and need to be cleaned. Fridge is open to dump through V3, V4, V1 and fridge valves. 1K Pot needle valve is open. He gas has been let into the cryostat vacuum jacket. Also some He gas in the IVC.
0910 Close all valves to He recovery
0915 G1 = 733 mbar, MC(Cernox) = 205 K (20 mW), Sorb = 250 K (max)
0920 MC is still warming slowly, extrapolating it looks like it won't hit 300 K until late tonight
0925 Close V4 to isolate the dump. Turn on 3He pump. Open V9 to connect pump outlet to dump. Open V6 to 13% to pump mix from fridge to dump.
0935 G1 down to 43 mbar, open V6 further
1015 Let about 680 mbar N2 gas into IVC. This is the pressure according to the mechanical gauge on the pump manifold. The Pfeiffer gauge reads 400 mbar. Note that the high-pressure mode on the Pfeiffer gauge reads too high when it is reading helium, see Dec 18 for example. But it reads closer to correct when it's N2 gas, at least in the 250-300 mbar range. (Now it's reading too low.)
1020 G2 (dump) = 789 mbar, G1 (fridge) = 0.3 mbar, both pretty flat. P1 (fridge) = 0.016 mbar and creeping down, but low enough. Stop pumping out fridge. Close still gate valve, condenser valve, V1, V3, and V6. Close V9 then turn off 3He pump.
G2 (Dump) = 789.2 mbar, Pbarometric = 29.71 inHg, Dump pressure (adjusted) = 782.1 mbar
1030 Let long hose length (4 or 5 meters) of N2 gas into IVC vacuum jacket to improve conductivity, speed warming.
1035 Adjust LakeShore bridge settings to range 2 kOhm, excitation 632 uV, much more appropriate to the Cernox thermometer in this range. Reading is much more uniform, but somewhat lower than before, 193 K
1210 MC(Cernox) = 202.8 K. Currently rising about 6 K/hr. At that rate it will reach room temperature early tomorrow morning.
1225 Let 5 more long hose-lengths of N2 into cryostat vacuum jacket, for good measure
1330 MC(Cernox) = 210.2 K
1340 Open Trap 1 up to G1 volume. Pressure rises to >1 bar (!). Pump this out using scroll pump through V12a+speedi, V2, V7, V2a, V5a, valve to scroll pump.
1346 Investigate non-functioning TES channels (DIB3 PA, DIB4 PA). Now that the detector is warming up, we should be able to measure ~ MΩ resistance from one TES bias/return channel to another, unless a channel has both of it's detector wirebonds broken.
DIB + - R (Ω)
3 PA bias PA ret 84
3 PA bias PB bias 620k
3 PA bias PC bias 780k
4 PA bias PA ret 79
4 PA bias PB bias 1.06M
4 PA bias PC bias 1.83M
4 PA bias PD bias 19M
These results indicate that the broken channels do not have both bias and return wirebonds broken. It could be the case that a single bond is broken, but since they're shorted by the 20mΩ shunt, its not possible to tell this way. Furthermore, the PA bias/return values indicate the circuit is intact at least up to the shunt resistors (i.e. not a stripline break).
1400 Thermostat reading/setpoint=71.8F/73F, Lab In/humidity = 23.7C/40%
1400 Pumped LN2 Cold Trap 1 down to 3 mbar.
1405 Checked LN2 Cold Trap 2: about 13 mbar in it. Pump this one out too.
1425 Put LN2 cold traps back in bath and fill with LN2 in preparation for cleaning mixture.
1700 MC(Cernox) = 228.4 K. IVC = 630 mbar (Pfeiffer). The trend looks like maybe it will be 285 K (around 50 F) by morning.

Next Day