| Status | Began warmup yesterday. Mixture has been pumped out of LN2 traps, which are currently warm and need to be cleaned. Fridge is open to dump through V3, V4, V1 and fridge valves. 1K Pot needle valve is open. He gas has been let into the cryostat vacuum jacket. Also some He gas in the IVC. |
| 0910 | Close all valves to He recovery |
| 0915 | G1 = 733 mbar, MC(Cernox) = 205 K (20 mW), Sorb = 250 K (max) |
| 0920 | MC is still warming slowly, extrapolating it looks like it won't hit 300 K until late tonight |
| 0925 | Close V4 to isolate the dump. Turn on 3He pump. Open V9 to connect pump outlet to dump. Open V6 to 13% to pump mix from fridge to dump. |
| 0935 | G1 down to 43 mbar, open V6 further |
| 1015 | Let about 680 mbar N2 gas into IVC. This is the pressure according to the mechanical gauge on the pump manifold. The Pfeiffer gauge reads 400 mbar. Note that the high-pressure mode on the Pfeiffer gauge reads too high when it is reading helium, see Dec 18 for example. But it reads closer to correct when it's N2 gas, at least in the 250-300 mbar range. (Now it's reading too low.) |
| 1020 | G2 (dump) = 789 mbar, G1 (fridge) = 0.3 mbar, both pretty flat. P1 (fridge) = 0.016 mbar and creeping down, but low enough. Stop pumping out fridge. Close still gate valve, condenser valve, V1, V3, and V6. Close V9 then turn off 3He pump.
G2 (Dump) = 789.2 mbar, Pbarometric = 29.71 inHg, Dump pressure (adjusted) = 782.1 mbar |
| 1030 | Let long hose length (4 or 5 meters) of N2 gas into IVC vacuum jacket to improve conductivity, speed warming. |
| 1035 | Adjust LakeShore bridge settings to range 2 kOhm, excitation 632 uV, much more appropriate to the Cernox thermometer in this range. Reading is much more uniform, but somewhat lower than before, 193 K |
| 1210 | MC(Cernox) = 202.8 K. Currently rising about 6 K/hr. At that rate it will reach room temperature early tomorrow morning. |
| 1225 | Let 5 more long hose-lengths of N2 into cryostat vacuum jacket, for good measure |
| 1330 | MC(Cernox) = 210.2 K |
| 1340 | Open Trap 1 up to G1 volume. Pressure rises to >1 bar (!). Pump this out using scroll pump through V12a+speedi, V2, V7, V2a, V5a, valve to scroll pump. |