| 0745 | Close V6 to stop circulation and allow mix to expand into still line |
| 0800 | Main bath = 0.2% |
| 0800 | Circulation pressures have fallen to 350 mbar. Still pressure reads 1000 mbar i.e. off scale. |
| 0800 | HV = 51.81 V, Vtest = 4.45 mV → Vdetector = 50.02 V.
Close vent valve on LN2 Dewar, which turns off the ebox purge. No noticeable decrease in LN supply according to the gauge. |
| 0805 | Close 1K Pot speedivalve to stop pumping on pot. Leaving needle valve open. Close V4a and valve to 4He pump. Open 4He outlet to “exhaust” and close recovery valve. Turn off 4He pump. |
| 0810 | Pump Trap 1 contents into dump. First close V1 to isolate fridge, then close V13a, open V9, open V5 to pump Trap 1 to dump.
When I opened V9, G2 rose from 255 mbar to 275 mbar. So there was definitely some mixture in the dump, must have been an overpressure release. |
| 0810 | Main bath = -0.6% |
| 0815 | Dial HV down to zero, short leads. Turn off HV power supply. |
| 0815 | 1K Pot has risen to 6K. Tower steady at 2K. |
| 0825 | G1 has bottomed out at ~0.4 mbar, LN2 Trap 1 is empty. Stop pumping trap out by closing V12a+speedi and V5. |
| 0830 | Check Trap 2 by opening V12b plus speedivalve. Strangely G1 jumps to 120 mbar when V12b opened but then immediately falls to 1.5 mbar. Pump this to dump through V5. |
| 0835 | G1 has bottomed out at ~0.4 mbar, LN2 Trap 2 is empty. Stop pumping trap out by closing V12b+speedi and V5. Then turn off 3He pump. Close V9 to isolate pump. |
| 0840 | Open V3, V1, and V4 to let mixture from fridge expand into dump. G1/G2 = 280 mbar |
| 0855 | Remove LN2 cold trap 1 and trap 2 from bath to let warm up |
| 1040 | Tower temperature has been pretty flat, currently 2.5 K. 1K Pot is maxed, Sorb is around 10 K. |
| 1040 | Main bath = -1.0% (minimum) |
| 1045 | Turn on MC heater to 20 mW. MC(Cernox) = 2.78 K |
| 1050 | Stop programs triggerfe2, towerfe3_01, dcrc_driver01. Disconnect power from DCRC1 and DCRC2. (DCRC4 which is connected to the NaI detector is still on.) |
| 1050 | MC temperature increasing, G2 increasing as mixture boils off, 1K Pot and Sorb temperatures falling as cold mixture blows past them. |
| 1100 | MC(Cernox) = 3.2 K and flattish, G1 (fridge/dump) = 375 mbar and rising, 1K Pot/Sorb = 4.78/6.1 K and flattish |
| 1105 | Turbo pump pressure = 1.4 nbar - this has been pumping on the IVC. Close pump valve. After 10 minutes IVC pressure reads 14 nbar. This is lower than last run - we weren't pumping on the IVC constantly last run. |
| 1125 | Let 2 tube-lengths of He gas into IVC. IVC → 0.1 mbar. G1 pressure increases rapidly, to 550 mbar. 1K Pot and Sorb temperatures fall a little, to 4.2/5.4 K. MC(Cernox) rises a little, → 3.6 K |
| 1205 | G1 = 655 mbar, MC(Cernox) = 4.75 K (20 mW), 1KPot/Sorb = 5.3/6.8 K, all basically flat. |
| 1445 | Remove PuBe source from lab, return it to PAN 442 |
| 1500 | Begin disassembling NaI source holder / south wall of shield |
| 1600 | NaI holder wall has been removed and lead/poly shield is mostly disassembled. |
| 1605 | G1 = 670 mbar, MC(Cernox) = 5.25 K (20 mW), 1KPot/Sorb = 6.7/8.7 K |
| 1615 | Let a 2“ length of hose worth of He gas into the cryostat vacuum jacket. (There was more He gas in the connector itself.) Recovery speed goes up a little.
MC, sorb and 1K Pot temperatures start warming more quickly |
| 1635 | Metal recovery hose is frosty but plastic tube isn't. |
| 1605 | G1 = 690 mbar, MC(Cernox) = 11.3↑ K (20 mW), 1KPot/Sorb = 6.999(max)/24.4 K |