start › cdms › k100 › run_diary › 20170518

2017/05/18 (Thursday)

Previous Day

Setup: DIB1/3/4 → 3.2 MHz filters 3/1/2 → cable 3/1/2 →gender changer 3/1/2 → HVConn 1/3/2 → HV 1/2 → DCRC55/DCRC33
PuBe source in paraffin barrel behind shield (west side)
Na-22 source hanging along southeast column of fridge stand
Took NaI-triggered data overnight @~50V, last sequence began at 07:58, looks OK.
0935 Tower = 32.9 mK, Main bath = 68.8%
0935 HV = 62.20 V, Vtest = 5.44 mV → Vdetector = 60.0 V(!) Apparently leakage decreased significantly overnight.
This is from the 8600A non-portable multimeter. There's a BNC cable that splits to two clamp probes and we have pins in the clamps stuck in the two test holes. Reseating these pins caused the Vtest reading to increase to something like 40 mV but then over several minutes it relaxed back down to <6 mV.
0945 Start 1 1-hour data series of 1-sided 50V HV coincidence data with Frontend Gain = (15/15/10)/(3/4/4), OG = 4/2, QET biases = table from Apr 28, with 1000 beginning randoms, no flashing, and a 20ADC bin threshold trigger on DCRC1 pA (internal trigger on Side1 pE). Running OR trigger of NaI detectors with PuBe source on west side of shield. NaI detector C has 311 mV threshold and detector B has 288 mV threshold
1120 Fill LN2 cold trap bath
1125 Dial down HV in steps and measure Vtest, waiting 10 minutes at each step before recording Vtest.
HV [V] Vtest [mV] Ileak [nA] Vloss [V ] Vdetector [V] Rtot [GOhm] Rleak [GOhm] Gleak [nS]
62.2 4.62 4.62 1.85 60.35 13.45 13.05 0.077
60 4.3 4.3 1.73 58.27 13.94 13.54 0.074
50 3.07 3.07 1.23 48.77 16.27 15.87 0.063
40.01 1.97 1.97 0.79 39.22 20.29 19.89 0.05
30 1.12 1.12 0.45 29.55 26.76 26.36 0.038
20 0.44 0.44 0.18 19.82 45.41 45.01 0.022
10 0.03 0.03 0.01 9.99 333 332.6 0.003

At HV=50V the leakage is similar to what it had been last weak. The leakage in nS is not constant but rather proportional to HV above 20V, with a dropoff below 20V. Not sure what that means, but it clearly isn't ohmic.

1230 Dial HV down to 0V
1240 Start 60 minute bake with DCRCs 1&2
1522 Take IbIs test data series. The sequence starts with ODBSET “/Logger/Write as CDMS series” n so it won't be copied to /rawdata with standard naming conventions. It should stay in /livedata as MIDAS runs 6573 and 6574. For processing purposes I created a dummy directory in /rawdata, 07170518_2500 (today's data but with a fictional time of day) and copied the two MIDAS series into it and renamed them with the standard dump naming conventions. (I did this as sudo cdmsssoft which seems to have given the directories and files the correct group and permissions.)
1556 Dial HV up to 65 V. After 10 minutes dial HV down to 50.00 V.
1630 DCRC1pC pulses are “hairy” (much larger high-frequency noise near the peak of a pulse). It looks like it only affects TES-saturated pulses. Adjusting front-end gain doesn't help much.
1630 HV = 50.00 V, Vtest (Fluke 8600a) = 3.11 mV → Vdetector = 48.8 V.
1633 Start 16 1-hour data series of 1-sided 50V HV coincidence data with Frontend Gain = (15/15/5)/(3/4/4), OG = 4/2, QET biases = table from Apr 28, with 1000 beginning randoms, no flashing, and a 500ADC bin threshold trigger on DCRC4 pA (external trigger). Running OR trigger of NaI detectors with PuBe source on west side of shield. NaI detector C has 311 mV threshold and detector B has 288 mV threshold
1655 Fill LN2 cold trap bath