2017/05/10 (Wednesday)

Previous Day

Setup: DIB1/3/4 → 3.2 MHz filters 3/1/2 → cable 3/1/2 →gender changer 3/1/2 → HVConn 1/3/2 → HV 1/2 → DCRC55/DCRC33
PuBe source in paraffin barrel behind shield (west side)
Took NaI-triggered data overnight @50V, looks good
0920 Tower/MC/Coldplate = 33.1/30.8/197 mK, Still = 1.3 K / 0.110 mbar, G1/G2 = 235/220 mbar, 1K Pot = 1.688 K / 28 mbar. Main Bath = 69.8%.
0925 HV = 50.01 V, Vtest = 4.7 mV → Vdetector = 47.9 V
0930 Fill LN2 cold trap bath.
1004 Start 1 1-hour data series of 1-sided 50V HV coincidence data with Frontend Gain = (15/15/10)/(4/5/4), OG = 4/2, QET biases = table from Apr 28, with 1000 beginning randoms, no flashing, and a 6ADC bin threshold trigger on DCRC1 pA (Side1 pE). Running OR trigger of NaI detectors with PuBe source on west side of shield. NaI detector C has 311 mV threshold and detector B has 288 mV threshold
1106 Dial HV down to 0 V.
1130 Begin 30 minute detector bake with DCRCs 1&2
1250 Dial HV up to 65 V. After 5 minutes dial down to 50.00 V.
1310 HV = 50.00 V, Vtest = 2.3 mV → Vdetector = 49.0 V
Tower = 33.2 mK, close to base
1316 Start 1 1-hour data series of 1-sided 50V HV coincidence data with Frontend Gain = (15/15/10)/(4/5/4), OG = 4/2, QET biases = table from Apr 28, with 1000 beginning randoms, no flashing, and a 6ADC bin threshold trigger on DCRC1 pA (Side1 pE). Running OR trigger of NaI detectors with PuBe source on west side of shield. NaI detector C has 311 mV threshold and detector B has 288 mV threshold
1426 Start 19 1-hour data series of 1-sided 50V HV coincidence data with Frontend Gain = (15/15/10)/(4/5/4), OG = 4/2, QET biases = table from Apr 28, with 1000 beginning randoms, no flashing, and a 1500ADC bin threshold trigger on DCRC4 pA (TTL signal). Running OR trigger of NaI detectors with PuBe source on west side of shield. NaI detector C has 311 mV threshold and detector B has 288 mV threshold
1700 Fill LN2 cold trap bath