2017/05/08 (Monday)
| Setup: DIB1/3/4 → 3.2 MHz filters 3/1/2 → cable 3/1/2 →gender changer 3/1/2 → HVConn 1/3/2 → HV 1/2 → DCRC55/DCRC33 | |
| PuBe source in paraffin barrel behind shield (west side) | |
| Took NaI-triggered data overnight @50V, looks good, although DCRC2 pD has significant lowish frequency noise, has been that way for a few days | |
| 0850 | Tower/MC/Coldplate = 33.0/30.6/189 mK, Still = 1.2 K / 0.110 mbar, G1/G2 = 235/215 mbar, 1K Pot = 1.690 K / 28 mbar. Main Bath = 71.8%. |
| 0850 | HV = 50.01 V, Vtest = 1.6 mV → Vdetector = 49.3 V |
| 0855 | Fill LN2 cold trap bath. |
| 0948 | Start 1 1-hour data series of 1-sided 50V HV coincidence data with Frontend Gain = (15/15/4)/(3/5/2), OG = 4/2, QET biases = table from Apr 28, with 1000 beginning randoms, no flashing, and a 20ADC bin threshold trigger on DCRC1 pA (Side1 pE). Running OR trigger of NaI detectors with PuBe source on west side of shield. NaI detector C has 311 mV threshold and detector B has 288 mV threshold |
| Note that I switched to triggering on a more stable channel. | |
| 0952 | Reduce trigger threshold on DCRC1pA from 20 to 18 |
| 0956 | Reduce trigger threshold on DCRC1pA from 18 to 16 |
| 1004 | Reduce trigger threshold on DCRC1pA from 16 to 14 |
| 1008 | Reduce trigger threshold on DCRC1pA from 14 to 12 |
| 1008 | Reduce trigger threshold on DCRC1pA from 12 to 10 |
| 1015 | Begin filling HPGe detector LN2 dewar. |
| 1021 | Reduce trigger threshold on DCRC1pA from 10 to 8 |
| 1024 | Reduce trigger threshold on DCRC1pA from 8 to 6 |
| 1028 | Reduce trigger threshold on DCRC1pA from 6 to 4 This was ok up to Dump 39, but at Dump 40 and later it looks like it's triggering too randomly. Maybe it got extra noisy at some point. So 6 seems safe but 4 maybe too low. |
| 1048 | Reduce trigger threshold on DCRC1pA from 4 to 6 |
| 1050 | Dial HV down to 0 V |
| 1055 | Begin 30 minute bake on DCRCs 1&2 |
| 1205 | Tower has cooled to 34.1 mK. Dial HV up to 65 V. |
| 1230 | Reduce HV to 50 V |
| 1240 | Looking at the DCRC2 channels, looks like pC and pD may both be showing some signs of 10 kHz oscillations, seem to be less noisy if QET bias is increased by 1 or 2 uA. But for now will leave them, maybe look more tomorrow. |
| 1245 | HV = 50.00 V, Vtest 0.9 mV → Vdetector = 49.6 V |
| 1249 | Start 1 1-hour data series of 1-sided 50V HV coincidence data with Frontend Gain = (15/15/4)/(3/5/2), OG = 4/2, QET biases = table from Apr 28, with 1000 beginning randoms, no flashing, and a 6ADC bin threshold trigger on DCRC1 pA (Side1 pE). Running OR trigger of NaI detectors with PuBe source on west side of shield. NaI detector C has 311 mV threshold and detector B has 288 mV threshold |
| 1353 | Start 19 1-hour data series of 1-sided 50V HV coincidence data with Frontend Gain = (15/15/4)/(3/5/2), OG = 4/2, QET biases = table from Apr 28, with 1000 beginning randoms, no flashing, and a 1500ADC bin threshold trigger on DCRC4 pA (TTL signal). Running OR trigger of NaI detectors with PuBe source on west side of shield. NaI detector C has 311 mV threshold and detector B has 288 mV threshold |
| 1708 | Fill LN2 cold trap full. |