UMN CDMS Wiki archive
start › cdms › k100 › run_diary › 20170419

2017/04/19 (Wednesday)

Previous Day

Setup: DIB1/3/4 → 3.2 MHz filters 3/1/2 → cable 3/1/2 → gender changer 3/1/2 → DCRC55(1)/33(2)/X(4)
0930 Fill LN2 cold trap full.
1010 Tower/MC/Coldplate = 34.1/31.5/199 mK, Still = 1.4 K / 0.120 mbar, G1/G2 = 260/237 mbar, 1K Pot = 1.697 K / 30 mbar. Main Bath = 31.6%. Boil off rate: faster than slow mode (predicted level: 35.1%)
1035 Installing NaI detectors on south shield wall. Isolate IVC from turbo pump, turn off turbo pump in preparation to move it.
1040 Power down DCRCs
1100 Modify HV board to bridge ground planes spanned by “crossover” channel, S1PA.
1140 Start LHe transfer (MB = 29.6%)
1146 Set up HV mode Setup: DIB1/3/4 → 3.2 MHz filters 3/1/2 → cable 3/1/2 →gender changer 3/1/2 → HVConn 1/3/2 → HV 1/2 → DCRC55/DCRC33
1156 Power DCRCs 55,33
1206 Restore DCRC1,2 phonon HV settings by inputting those present in the last saved HV .odb file (livedata/run06056.odb)
1209 Zap SQUIDs
1240 End LHe transfer (MB = 91.9%)
1334 Pulses seen on all 6 channels, DCRC2 PB looks a little noisy, but can definitely see pulses on it now!
1425 Finish IbIs on DCRC2pB
1530 Attach south shield frame with NaI detectors attached.
1622 Re-attach turbo pump manifold to IVC line. Rough pump on pump manifold and start up turbo pump while isolated.
1704 Rough pump turbo manifold and line up to IVC manual valves which are still closed. Begin turbo pumping on line up to IVC manual valves. Let's leave this pumping over night with IVC isolated.
1717 Prebias HV = 65 V.
1715 From April 11 it seems that the current QET biases give Ib*prefactor=40. This may be too high, but for now, set DCRC2pB's to be similar for consistency's sake. Based on IbIs, DCRC2pB should have QET bias = -121.17 uA to achieve Ib*pf = 40
1725 Reduce HV to 50V
1727 Start sequence of 12, 1-hour data series of 1-sided 50V data with Frontend Gain = (20/20/10/-)/(-/3/4/2), OG = 4/2, QET biases = described above at 1715, with 1000 beginning randoms, no flashing, and a 17ADC bin threshold trigger on DCRC1 pA
1730 HV = 49.99 V, V_leak = 2.8 mV, V_crystal = 48.76 V
1737 Turbo pressure is 4.0e-6 mbar
1740 Fill LN2 cold trap full
1743 Stop the run prematurely, DCRC1pC has some HF noise at the top of pulses, DCRC2PB and DCRC2pC are railed.
1747 Reduce DCRC1pC FE gain to 5. This helps a little bit. Probably it needs to be zapped though. Adjust other channel driver offsets.
1751 Start sequence of 12, 1-hour data series of 1-sided 50V data with Frontend Gain = (20/20/10/-)/(-/3/4/2), OG = 4/2, QET biases = described above at 1715, with 1000 beginning randoms, no flashing, and a 17ADC bin threshold trigger on DCRC1 pA

Next Day