Setup: DIB1/3/4 → 3.2 MHz filters 3/1/2 → cable 3/1/2 → gender changer 3/1/2 → DCRC55(1)/33(2)/X(4)
0910 Tower = 33.6 mK, main bath = 60.4%. Boil off rate: close to slow mode (predicted level: 61.9%)
0915 Trying to find good QET bias settings for all 9 working channels.
Notice that DCRC1pC is falling into s/c before getting down to yesterday's setting, so will need to redo IbIs on that one. Also still need to do DCRC1pB
0930 Finish IbIs on DCRC1pB/Side1pB.
1000 Finish IbIs on DCRC1pC/Side1pD
Results are quite different from yesterday's measurement. Not sure why. Yesterday's was shortly after the transfer ended, not sure if that should matter.
1010 Fill LN2 cold trap bath.
1035 DCRC4pB also seems to be behaving differently today. Redo its IbIs too.
1055 Finish IbIs on DCRC4pB/Side2pF. It's very different from Friday's measurement
QET bias settings to achieve Ib*prefactor = 20 uA
DCRC channel Detector channel Ib*prefac [uA] QET bias date measured
DCRC1 pA Side1 pE 20 124.5 April 15
DCRC1 pB Side1 pB 20 137.1 April 17
DCRC1 pC Side1 pD 20 123.5 April 17
DCRC1 pD Side2 pE 20 135.5 April 14
DCRC2 pA Side2 pB not working
DCRC2 pB Side2 pA 20 -123.7 April 14
DCRC2 pC Side1 pA 20 -121.7 April 14
DCRC2 pD n/c not connected
DCRC4 pA Side2 pC not working
DCRC4 pB Side2 pF 20 142.3 April 17
DCRC4 pC Side1 pC 20 141.9 April 14
DCRC4 pD Side1 pF 20 150.9 April 16
Measurements are from most recent IbIs in current setup
1112 Start sequence of 20 4, 1-hour data series of 2-sided 0V data with Frontend Gain = (20/20/15/20)/(-/4/4/-)/(-/3/5/5), OG = 4/2/4, QET biases = table above, with 1000 beginning randoms, no flashing, and a 15ADC bin threshold trigger on DCRC1 pA
1116 Adjust trigger threshold up from 15 to 17 ADC bins
1116 Adjust trigger threshold up from 17 to 20 ADC bins
1248 Adjust trigger threshold down from 20 to 18 ADC bins
1314 Adjust trigger threshold down from 18 to 17 ADC bins
1430 Stop data early to do the LHe transfer.
1430 Start LHe transfer (MB = 52.7%)
1440 Tower/MC/Coldplate = 33.1/30.8/197 mK, Still = 1.3 K / 0.110 mbar, G1/G2 = 253/227 mbar, 1K Pot = 1.680 K / 28 mbar.
1535 Finish LHe transfer (MB = 90.7%)
1548 Start sequence of 17, 1-hour data series of 2-sided 0V data with Frontend Gain = (20/20/15/20)/(-/4/4/-)/(-/3/5/5), OG = 4/2/4, QET biases = table above, with 1000 beginning randoms, no flashing, and a 17ADC bin threshold trigger on DCRC1 pA
1550 Tower = 33.4 mK
1830 Fill LN2 cold trap