start › cdms › k100 › run_diary › 20170416

2017/04/16 (Sunday)

Previous Day

Setup: DIB1/3/4 → 3.2 MHz filters 3/1/2 → cable 3/1/2 → gender changer 3/1/2 → DCRC55(1)/33(2)/X(4)
1345 Fill LN2 cold trap bath.
1345 Tower/MC/Coldplate = 33.4/31.0/193 mK, Still = 1.2 K / 0.120 mbar, G1/G2 = 247/226 mbar, 1K Pot = 1.698 K / 30 mbar. MB = 29.8%. Boil off rate: faster than slow mode (predicted level: 32.6%).
1400 Redo IbIs on DCRC4pD/DIB4pD/Side1pF
1500 Start LHe transfer (MB = 28.3%)
1610 End LHe transfer (MB = 82.0%, partial fill - dewar ran empty)
1650 Finish IbIs on DCRC1pC/DIB1pC/Side1pD
QET bias settings to achieve Ib*prefactor = 20 uA
DCRC channel Detector channel Ib*prefac [uA] QET bias date measured
DCRC1 pA Side1 pE 20 124.5 April 15
DCRC1 pB Side1 pB 20 139.7 March 31*
DCRC1 pC Side1 pD 20 121.1 April 16
DCRC1 pD Side2 pE 20 135.5 April 14
DCRC2 pA Side2 pB not working
DCRC2 pB Side2 pA 20 -123.7 April 14
DCRC2 pC Side1 pA 20 -121.7 April 14
DCRC2 pD n/c not connected
DCRC4 pA Side2 pC not working
DCRC4 pB Side2 pF 20 137.3 April 15
DCRC4 pC Side1 pC 20 141.9 April 14
DCRC4 pD Side1 pF 20 150.9 April 16
Measurements are from most recent IbIs in current setup

* this one is too old, with the HV setup. Should be redone
1924 Start test series, randoms only