2016/12/14 (Wednesday)

Previous Day

Setup: DIB1/3/4 → 3.2 MHz filters → cables 1,2,3 → HV board inputs 1,3,2 → HV board outputs 1,2 → DCRC55(1)/50(3)
NaI-coincident data taken overnight, HV=60V. The last series in the sequence ended at 7:25 (10 hours of data starting at 21:25 last night). BatViewer shows that all 5 channels are still good, nothing railed, reasonable pulse shapes.
1026 Do a 30 minute bake 30s on / 5s off, prepare for another 3 hours of data by pre-biasing to 75V for 5 min, this will end just after the PuBe source arrives
1140 Tower/MC/Coldplate = 33.3/34.5/203 mK, Still = 1.03 K / 110 μbar, G1/G2 = 192/171 mbar, 1K Pot = 1.68 K / 22 mbar. MB = 66.3%. Boil off rate: faster than slow mode.
1345 Fill LN2 cold trap full
1250 do a 6 minute bake
1354 PuBe source in parafin container next to fridge, outside of shielding, will return at 16:15
1400 overbias to 115V wait 5 minutes
1415 see pulses, modify source position to decrease rate in ZIP do ZAP
1423 source rate still a bit high – 130Hz
1425 moved it back about a meter and got to 90Hz
1615 removed PuBe source, Andy gave us the key (on Matt's desk) we can use it anytime between now and Dec 31st 8am-4pm, no weekends, in the paraffin case only
1627 ramp voltage down and do a flash, follow with bkgnd running in same conditions as above
1659 the rate post-PuBe source seems to be around 36 Hz, so we didn't pick up a lot of activation from this
1702 note DCRC1 triggering is enabled in all the data from today and yesterday, I was worried about this because in the past DCRC1 has produced spurious triggers, but I just checked with the “rt” command to the telnet session and it is apparently not issuing spurious triggers, maybe it's OK now. DCRC3 is also not producing spurious triggers