2016/09/21 (Wednesday)

Previous Day

DCRC1 and DCRC3 powered on and connected through HV setup EBoxDIB 2/3 → 3.2 MHz filters → Cables → remapping board I/remapping board II → HVside2board “DIB1”/“DIB2” → DCRC55(1)/50(3). Remapping boards have all possible lines switched off. .
Baked using UV LEDs for 9 hours last night. Tower was at 0.4 K until 07:17 then fell to base over about 40 minutes.
0835 Tower/MC/Coldplate = 93/58/??(1.70 kΩ) mK, Still = 1.13 K / 85 μbar, G1/G2 = 160/150 mbar, 1K Pot = 1.51 K / 3.4 mbar. MB = 36.5%
The thermometers connected to “FRIDGE OUT 1”/“SENSOR 1” cable to the thermometer breakout box seem to be reading wrong. Coldplate is off scale, Still is higher than usual.
Boiloff rate: good
1K Pot has been pretty stable.
0845 Fill LN2 cold trap bath.
1015 Switch HV Switchboards and Flash&Flash python script back to using IR LEDs. Not flashing until after the first data series.
1022 Manually start 20min background data series at HV = 10V, DCRC1/Readout= -10V with FG=5, OG=8, with 1000 initial randoms, 1000 ending randoms with 150 ADC Bin trigger threshold on DIB1 Qo.
1043 Cease data run. Ground detector, flash with IR LEDs for 30s, then cool down for 10 minutes.
1058 Manually start 20min background data series at HV = 10V, DCRC1/Readout= -10V with FG=5, OG=8, with 1000 initial randoms, 1000 ending randoms with 150 ADC Bin trigger threshold on DIB1 Qo.
1119 Cease data run. Ground detector, flash with IR LEDs for 30s, then cool down for 10 minutes.
1135 Manually start 20min background data series at HV = 10V, DCRC1/Readout= -10V with FG=5, OG=8, with 1000 initial randoms, 1000 ending randoms with 150 ADC Bin trigger threshold on DIB1 Qo.
1156 Cease data run. Ground detector. Remove resistor from breadboard circuit (thus breaking the external circuit), and flip Qo-Feedback switch to ON. Apply -10V on Qo using DCRC1, and check pulses to see if threshold needs to be adjusted, but 150ADC Bins seems to trigger pretty well.
1209 Zero charge bias, and flash with IR LEDs for 30s, then cool down for 10 minutes.
1225 Check the pulses again, and the threshold looks fine. I've realized I forgot to turn off the logger for the past two times I checked the pulses with the SQUID GUI, so there will be some short runs that are just tests.
1229 Manually start 20min background data series at HV = 10V, DCRC1/Readout= -10V with FG=5, OG=8, with 1000 initial randoms, 1000 ending randoms with 150 ADC Bin trigger threshold on DIB1 Qo.
1249 Cease data run. Ground detector, flash with IR LEDs for 30s, then cool down for 10 minutes.
1307 Manually start 20min background data series at HV = 10V, DCRC1/Readout= -10V with FG=5, OG=8, with 1000 initial randoms, 1000 ending randoms with 150 ADC Bin trigger threshold on DIB1 Qo.
1329 Cease data run. Ground detector, flash with IR LEDs for 30s, then cool down for 10 minutes.
1346 Manually start 20min background data series at HV = 10V, DCRC1/Readout= -10V with FG=5, OG=8, with 1000 initial randoms, 1000 ending randoms with 150 ADC Bin trigger threshold on DIB1 Qo.
1406 Cease data run. Ground detector, and bake for 90 minutes with the IR LEDs during the transfer.
1500 Commence LHe transfer (26.8%).
1610 Cease LHe transfer (94.5%).
1615 Replace the resistor in the breadboard circuit (thus breaking the external circuit), and flip Qo-Feedback switch to ON. Apply -10V on Qo using DCRC1, and check pulses. It looks in working order.
1620 Let pressure out of the Main Bath.
1637 Now that the fridge is cool enough (Tower = 85mK), manually start 20min background data series at HV = 12V, DCRC1/Readout= -12V with FG=5, OG=8, with 1000 initial randoms, 1000 ending randoms with 150 ADC Bin trigger threshold on DIB1 Qo.
1640 Let pressure out of the Main Bath.
1658 Cease data run. Ground detector, flash with IR LEDs for 30s, then cool down for 10 minutes.
1713 Manually start 20min background data series at HV = -12V, DCRC1/Readout= 12V with FG=5, OG=8, with 1000 initial randoms, 1000 ending randoms with 150 ADC Bin trigger threshold on DIB1 Qo.
1735 Cease data run. Ground detector, flash with IR LEDs for 30s, then cool down for 10 minutes. Release pressure from Main Bath.
1739 Fill LN2 cold trap full.
1749 Manually start 20min background data series at HV = 12V, DCRC1/Readout= -12V with FG=5, OG=8, with 1000 initial randoms, 1000 ending randoms with 150 ADC Bin trigger threshold on DIB1 Qo.
1809 Cease data run. Ground detector, bake with IR LEDs for 12 hours.