2016/04/07 (Thursday)
| Tower installation | |
|---|---|
| 1055 | Scotch-brite the top surface of the adapter plate on the tower again |
| 1100 | Scotch-brite bottom surface of 30mK can |
| 1110 | Begin tower installation |
| 1145 | Tower adapter plate attached and all screws torqued into 30mK can. Attach thermometer, and check that the connection is made. Lakeshore bridge should have the Tower Thermometer reading ~1kΩ. Currently reading 1.0001kΩ. Looking good. |
| 1645 | Striplines are plugged in. Performed room temperature electrical checkout at breakout box, all looks good. |
| 1710 | After sealing IVC, did another quick electrical check at breakout box. Sadly, DIB1pB Squid Bias-Return is now reading open. Will need to investigate further. |
|---|
| Soudan Tower IT4 | |
|---|---|
| 1125 | Inspect 4K base of tower around pin sockets. No debris seen. |
| 1130 | Check for shorts to chassis on all sockets at 4K. None found. |
| IT4 Side Coaxes | ||
|---|---|---|
| Face | Length | ID |
| 1 | Short (position 1) | ST-3-10 |
| 2 | Medium (position 2) | ST-4-9 |
| 3 | Long (position 3) | ST-5-11 |
| 4 | Short (position 1) | ST-3-9 |
| 5 | Medium (position 2) | ST-4-10 |
| 6 | Long (position 3) | ST-5-10 |
| 1150 | Remove side coax from Face 1 of IT4. To do so first remove two round-head screws from tower side and two flat-head screws on detector side. Covered the pins with protector boards. However we won't have enough protector boards for all these side coaxes. |
|---|---|
| 1155 | Put IT4 back in purge cabinet. |