2016/04/06 (Wednesday)

Previous Day

1045 Go into clean room and begin assembling tower with G114 using Faces 2,5, and newly repaired Face 3 of tower.
1425 Bring Tower Shipping Vessel containing T4 from Soudan into cleanroom.
1430 Perform tower level electrical check.
Detector G114 DIB 1
side coax ZB1-4-4 date/who: 2016-04-06 DS NM
face 5
+lead - lead signal typical measured
1 2 QET A 100 to 200 Ω 834 Ω (Doing 2→1 gives 1.5kΩ)
2 4 QETA-B 100 to 200 Ω 275 Ω
3 4 QET B 100 to 200 Ω 254 Ω
4 6 QETB-C 100 to 200 Ω 253 Ω
5 6 QET C 100 to 200 Ω 221 Ω
6 8 QETC-D 100 to 200 Ω 216 Ω
7 8 QET D 100 to 200 Ω 166 Ω
2 8 QETA-D 100 to 200 Ω 226 Ω
chassis 9 LED1 Rforward 8 to 9 MΩ open
chassis 9 LED1 Vforward 1.0 to 1.1 V open
9 chassis LED1 Rreverse <1 nS <1 nS
chassis 10 LED2 Rforward 8 to 9 MΩ 690K
chassis 10 LED2 Vforward 1.0 to 1.1 V 1.086 V
10 chassis LED2 Rreverse <1 nS <0.3 nS
1 11 Qo Rbias 24 to 26 nS 26 nS
12 11 Qo C coupling <1 nS <0.5 nS
12 13 Qo Rfeedback 24 to 26 nS 26 nS
1 14 Qi Rbias 24 to 26 nS 26 nS
15 14 Qi C coupling <1 nS <2 nS
15 16 Qi Rfeedback 24 to 26 nS 26 nS
Detector G114 DIB 2
side coax ZB1-4-6 date/who: 2016-04-06 DS NM
face 2
+lead - lead signal typical measured
1 2 QET A 100 to 200 Ω 246 Ω
2 4 QETA-B 100 to 200 Ω 211 Ω
3 4 QET B 100 to 200 Ω 131 Ω
4 6 QETB-C 100 to 200 Ω 218 Ω
5 6 QET C 100 to 200 Ω 202 Ω
6 8 QETC-D 100 to 200 Ω 199 Ω
7 8 QET D 100 to 200 Ω 131 Ω
2 8 QETA-D 100 to 200 Ω 225 Ω
chassis 9 LED1 Rforward 8 to 9 MΩ 690 kΩ
chassis 9 LED1 Vforward 1.0 to 1.1 V 1.090 V
9 chassis LED1 Rreverse <1 nS <0.5 nS
chassis 10 LED2 Rforward 8 to 9 MΩ 690 kΩ
chassis 10 LED2 Vforward 1.0 to 1.1 V 1.088 V
10 chassis LED2 Rreverse <1 nS <0.5 nS
1 11 Qo Rbias 24 to 26 nS 19 nS
12 11 Qo C coupling <1 nS <0.3 nS
12 13 Qo Rfeedback 24 to 26 nS 26 nS
1 14 Qi Rbias 24 to 26 nS 25 nS
15 14 Qi C coupling <1 nS <0.4 nS
15 16 Qi Rfeedback 24 to 26 nS 26 nS
Detector G114 DIB 3
side coax ZB1-4-8 date/who: 2016-04-06 DS NM
face 3
+lead - lead signal typical measured
1 2 QET A 100 to 200 Ω 117 Ω
2 4 QETA-B 100 to 200 Ω 188 Ω
3 4 QET B 100 to 200 Ω 163 Ω
4 6 QETB-C 100 to 200 Ω 151 Ω
5 6 QET C 100 to 200 Ω 114 Ω
6 8 QETC-D 100 to 200 Ω 186 Ω
7 8 QET D 100 to 200 Ω 166 Ω
2 8 QETA-D 100 to 200 Ω 188 Ω
chassis 9 LED1 Rforward 8 to 9 MΩ open
chassis 9 LED1 Vforward 1.0 to 1.1 V open
9 chassis LED1 Rreverse <1 nS open
chassis 10 LED2 Rforward 8 to 9 MΩ open
chassis 10 LED2 Vforward 1.0 to 1.1 V open
10 chassis LED2 Rreverse <1 nS open
1445 Everything looks good. Place tower in purge cabinet overnight.
1545 Begin opening Soudan IT4 shipping vessel in clean room
Pressure = 14.2 PSI. (Somewhat higher than tape mark, which makes sense since it was pressurized in the mine which has a higher atmospheric pressure.)
Remove small cap at end of pump port. Open valve, just enough to let pressure out slowly.
1600 Remove top of shipping vessel using 5/32 hex key
Cover disk at 4K stage is labeled ZB-8. I don't think this is the tower ID; the cold hardware elog has tower IDs that all start with ZB1-something.
update: I think this tower is ZB1-7.
1615 Remove screws attaching tower ears to vessel base using M4 hex key. Lift tower out of vessel.
1622 Place tower in large purge cabinet

Next Day