2016/02/23 (Tuesday)
| Setup: iZIP mode, DIB1/2/3 → cables → 3.2 MHz filters → DCRC55/33/50 | |
| Taking Y-Be data overnight. Last flash at ~6:45. | |
| 0900 | Tower/MC(H35)/MC sensor/Cold Plate = 31.7/28.1/41/156 mK, Still = 1.11 K / 0.085 mbar, G1/G2 = 120/110 mbar, 1K Pot = 1.50 K / 1.7 mbar, Main bath = 75.6% |
| 0900 | Remove Y source from fridge. |
| 0905 | Tweak 1K Pot needle valve open additional 5 degrees. |
| 0910 | Fill LN2 cold trap Dewar. |
| 1130 | Perform mini IbIs on DIB3 in standard setup. After that, unpower DCRC50, remove DSUB pi-filter from DIB3, attach pi-filter switchboard with all phonon related filters turned on, and repower DCRC50. Zap SQUIDs. (Only lines on pi-filter switchboard not turned on are 42, 26, 11, 27, 12, 30, 15, 31, 16, 50, 45, 46, 49) |
| 1220 | Perform mini IbIs on DIB3 in with pi-filter switchboard attached and all phonon lines filtered. After that, unpower DCRC50, switch off all phonon lines and switch on all charge lines on pi-filter switchboard, and repower DCRC50. Zap SQUIDs. (Only lines filtered on pi-filter switchboard are 11, 27, 12, 30, 15, 31, 16, 50) |
| 1250 | Perform mini IbIs on DIB3 in with pi-filter switchboard attached and all charge lines filtered. Detach pi-filter switchboard, reattach pi-filter DSUB, and repower DCRC50. Zap SQUIDs. |
| 1345 | Place Ba source inside shield near detector height. |
| 1350 | Set all phonon channels to phonon trace length of 1024 bins (2048 MIDASbins) and charge channels to a charge trace length of 512 bins (1024 MIDASbins). Set both phonon and charge prepulse lengths to 100 MIDASbins. |
| 1352 | Check phonon traces. DIB2 pA and DIB3 pC railed, but changing Driver Offset brings them back. |
| 1355 | Start 30min Ba series with shortened trace lengths from above at -4/+4 V 0/0 V with FG=5/2/5, OG=3, with 1000 initial randoms, 1000 ending randoms, with no flash beforehand. Trigger on DIB1 phA with threshold of 20ADC bins, with 150 us pileup cut window. |
| 1455 | Place Ba source back in safe. |
| 1500 | Unpower DCRC50, detach DSUB pi-filter, attach pi-filter switchboard with only pB phonon lines filtered, and repower DCRC50. Zap SQUIDs. |
| 1610 | Perform mini IbIs on DIB3 in with pi-filter switchboard attached and all phB phonon lines filtered. Detach pi-filter switchboard, reattach pi-filter DSUB, and repower DCRC50. Zap SQUIDs. |
| 1625 | Set all phonon channels to phonon trace length of 512 bins (1024 MIDASbins) and charge channels to a charge trace length of 256 bins (512 MIDASbins). Set both phonon and charge prepulse lengths to 100 MIDASbins. |
| 1630 | Check phonon traces. A few channels railed, changing Driver Offset brings them back. |
| 1635 | Place Ba source inside shield near detector height. |
| 1636 | Start 30min Ba series with doubly shortened trace lengths from above at -4/+4 V 0/0 V with FG=5/2/5, OG=3, with 1000 initial randoms, 1000 ending randoms, with no flash beforehand. Trigger on DIB1 phA with threshold of 20ADC bins, with 150 us pileup cut window. |
| 1710 | Set all phonon channels to phonon trace length of 8192 bins (16384 MIDASbins) and charge channels to a charge trace length of 16384 bins (32768 MIDASbins). Set both phonon and charge prepulse lengths to 1000 MIDASbins. |
| 1713 | Start 30min Ba series with elongated trace lengths from above at -4/+4 V 0/0 V with FG=5/2/5, OG=3, with 1000 initial randoms, 1000 ending randoms, with no flash beforehand. Trigger on DIB1 phA with threshold of 20ADC bins, with 150 us pileup cut window. |
| 1715 | Fill LN2 cold trap full. |
| 1745 | Set all phonon channels to phonon trace length of 4096 bins (8192 MIDASbins) and charge channels to a charge trace length of 8192 bins (16384 MIDASbins). Set both phonon and charge prepulse lengths to 500 MIDASbins. |
| 1753 | Start 14 hours of 30min Y-Be series at -4/+4 V with standard trace lengths with FG=5/2/5, OG=3, with 1000 initial randoms, 1000 ending randoms, with 30s flash and 25 minute cooldown in between series. Trigger on DIB1 phA with threshold of 20ADC bins and DIB1 qInner&qOuter each with threshold 210 ADC bins, with 150 us pileup cut window. |
| 1755 | Place strong 88-Y source in Be disc source holder. |
Spreadsheet of Mini-IbIs data