start › cdms › k100 › run_diary › 20160215

2016/02/15 (Monday)

Previous Day

Setup: iZIP mode, DIB1/2/3 → cables → 3.2 MHz filters → DCRC55/33/50
Taking Y-Be data overnight, last flash ~ 7:20. Biases accidentally left at 0 V for last night's data.
0905 Tower/MC(H35)/MC sensor/Cold Plate = 31.8/28.3/44/167 mK, Still = 1.26 K / 0.09 mbar, G1/G2 = 125/125 mbar, 1K Pot = 1.53 K / 2.6 mbar, Main bath = 69.6%
0905 Tweak 1K Pot needle valve open additional ~5-10 degrees.
0910 Fill LN2 cold trap Dewar.
1020 Check phonons, all seem OK
1024 Start test background series with standard setup at 0/0 V with FG=5/2/5, OG=3, with 1000 initial randoms, 0 ending randoms, no flash. Trigger on DIB1 phA with threshold of 20ADC bins and DIB1 qInner&qOuter each with threshold 210 ADC bins, with 150 us pileup cut window. Stop after initial randoms are done.
1035 Remove D-sub filter from DIB 3. First turned off MIDAS programs. Repowered DCRC and set up timing link. Restart MIDAS and zap SQUIDs.
1050 Check phonons, see 70 Hz sinusoid. Something wrong with timing link maybe, or stray test signal. Redoing timing link doesn't help.
1055 Power cycle DCRC50(DIB3).
1100 Sinusoid still present in DIB1 phonons. Try timing link once more. Doesn't help.
1105 Power cycle DCRC33(DIB2), which I accidentally power cycled earlier. Reestablish timing link. Sinusoid still present in DIB1 phonons.
1110 On advice from David (who is home sick today), did rd 3d on the DCRCs. Expect 09D7, indicating test signal off. This was true for DIB1 but not for DIB2 or DIB3, which read 0DD7, indicating test signal on. Did wr 3d 09D7 to both DIB2 and DIB3. However, sinusoid still present in DIB1 phonons.
1125 Power cycle all three boards. Start boards, find that 3d is 0DD7 on all three, change to 09D7. Reestablish timing link. Zap SQUIDs.
1140 Check phonons again, sinusoid still present, seen in DIB1 and DIB3 (all DIB2 railed).
1150 Don't know what to do. Powering down DCRCs for the time being.
1206 During current power cycling add Pi filter back on to Dib3 DCRC50, and power boards back on. Reestablish timing link and ZAP SQUIDs.
1230 No sinusoid seen on any phonon channels. Just to check, did wr 3d to all board telnet sessions. DCRC1 returned 09D7, DCRC2 and DCRC3 returned 0DD7.
1259 (Bad Trace lengths) Start test background series with standard setup at 0/0 V with FG=5/2/5, OG=3, with 1000 initial randoms, 0 ending randoms, no flash. Trigger on DIB1 phA with threshold of 20ADC bins and DIB1 qInner&qOuter each with threshold 210 ADC bins, with 150 us pileup cut window. Stop after initial randoms are done.
1305 Remove D-sub filter from DIB 3. First turned off MIDAS programs. Repowered DCRC. Restart MIDAS programs and zap SQUIDs. Set up timing link.
1320 DCRC1 PC railed. Changed offset from 0.12 V to -0.12 V. Otherwise all phonons look good.
1325 (Bad Trace lengths) Start test background series with DIB3 unfiltered at 0/0 V with FG=5/2/5, OG=3, with 1000 initial randoms, 0 ending randoms, no flash. Trigger on DIB1 phA with threshold of 20ADC bins and DIB1 qInner&qOuter each with threshold 210 ADC bins, with 150 us pileup cut window. Stop after initial randoms are done.
1330 Add FilterBoard(all filters off) to DIB3. First turned off MIDAS programs. Repowered DCRC. Restart MIDAS programs and zap SQUIDs. Set up timing link.
1352 DCRC3 PA railed. Changed offset to -0.15 V.
1406 Start test background series with pi-filter board on DIB3 with all switches off at 0/0 V with FG=5/2/5, OG=3, with 1000 initial randoms, 0 ending randoms, no flash. Trigger on DIB1 phA with threshold of 20ADC bins and DIB1 qInner&qOuter each with threshold 210 ADC bins, with 150 us pileup cut window. Stop after initial randoms are done.
1418 Switch on all FilterBoard lines. First turned off MIDAS programs. Repowered DCRC. Restart MIDAS programs and zap SQUIDs. Set up timing link.
1435 DCRC3 PA was railed. Tried correcting it by changing polarity to 0 (feedback disable) and then back to 1 (non-inverting). Could see no effect in Squid tuning. After a while, figured out that triggerfe2 had crashed. Restarted triggerfe2. Tried changing polarity again with no change in DCRC3 PA.
1450 Changed DCRC PA offset to +0.08 V
1452 Start test background series with pi-filter board on DIB3 with all switches on at 0/0 V with FG=5/2/5, OG=3, with 1000 initial randoms, 0 ending randoms, no flash. Trigger on DIB1 phA with threshold of 20ADC bins and DIB1 qInner&qOuter each with threshold 210 ADC bins, with 150 us pileup cut window. Stop after initial randoms are done.
1459 Replace Filterboard with standard Pi Filter on DIB3 (i.e. we're back in Standard Mode). First turned off MIDAS programs. Repowered DCRC. Set up timing link. Restart MIDAS programs and zap SQUIDs.
1645 1K Pot pressure down to 5.5 mbar (1.62 K), tweak needle valve open additional 5 degrees.
1655 Fill LN2 cold trap Dewar
1700 Start 15 hours of 30min Y-Be series at -4/+4 V with FG=5/2/5, OG=3, with 1000 initial randoms, 1000 ending randoms, with 30s flash and 25 minute cooldown in between series. Trigger on DIB1 phA with threshold of 20ADC bins and DIB1 qInner&qOuter each with threshold 210 ADC bins, with 150 us pileup cut window.
1725 Place Y source in shielding.

Next Day