2016/02/14 (Sunday)

Previous Day

Setup: iZIP mode, DIB1/2/3 → cables → 3.2 MHz filters → DCRC55/33/50
Taking Y-Be data overnight, last flash ~ 11:20
1320 Tower/MC(H35)/MC sensor/Cold Plate = 31.5/27.8/42/160 mK, Still = 1.14 K / 0.09 mbar, G1/G2 = 130/120 mbar, 1K Pot = 1.59 K / 4.5 mbar, Main bath = 33.1%
1330 Begin LHe transfer at 32.6%
1335 Fill LN2 cold trap Dewar.
1335 Tweak 1K Pot needle valve open additional ~5 degrees.
1410 1K Pot has fallen to 1.66 K / 10 mbar, tweak open a little more
1414 Finish LHe transfer at 90.4%
1417 Start 17.7 hours of 30min Y-Be series at -4/+4 V 0/0 V with FG=5/2/5, OG=3, with 1000 initial randoms, 1000 ending randoms, with 30s flash and 25 minute cooldown in between series. Trigger on DIB1 phA with threshold of 20ADC bins and DIB1 qInner&qOuter each with threshold 210 ADC bins, with 150 us pileup cut window.
dammit - forgot to set biases, took data at 0 bias
1420 Let pressure out of main bath. 1K Pot currently at 1.71 K / 19 mbar.