2016/02/10 (Wednesday)

Previous Day

Setup: iZIP mode, DIB1/2/3 → cables → 3.2 MHz filters → DCRC55/33/50
Cf activation overnight
0905 1K Pot pressure has fallen to 3.2 mbar. Tweak needle valve open additional 5 degrees.
0900 Tower/MC(H35)/MC sensor/Cold Plate = 31.5/27.8/42/159 mK, Still = 1.17 K / 0.080 mbar, G1/G2 = 120/110 mbar, 1K Pot = 1.49 K / 1.7 mbar, Main bath = 38.4%
0905 Tweak 1K Pot needle valve open additional 5 degrees. After 30 minutes settles to 1.7 K / 20 mbar, but falling slowly.
0905 Remove Cf source from fridge
0910 Fill LN2 cold trap Dewar.
0920 Check phonons:
DIB2pD railed: relock fixed it. Had to do this twice.
DIB1pB Showing no pulses, only pulseless traces. The QET bias was set to 127.2 uA - this seems to be too high now. Reduced QET bias to 80 uA, now strong pulses seen.
It is suspected that the processing is currently not working correctly because the corrected charge prepulse setting is putting the charge pulses outside of the processing window. So we'll take some test series like yesterday but with the charge prepulse setting returned to 250 as it was before Feb 4.
0940 Start 15min test background series with standard setup at -4/+4 V with FG=5/2/5, OG=3, with 1000 initial randoms, 0 ending randoms, no flash. Trigger on DIB1 phA with threshold of 20ADC bins with 150 us pileup cut window. Charge prepulse set to 250 on all three DCRCs.
0955 Stop run. Enable high-threshold on DIB1 Qi&Qo.
0957 Start 15min test background series with standard setup at -4/+4 V with FG=5/2/5, OG=3, with 1000 initial randoms, 0 ending randoms, no flash. Trigger on DIB1 phA with threshold of 20ADC bins and DIB1 qInner&qOuter each with threshold 210 ADC bins,, with 150 us pileup cut window.
1012 Stop run.
1555 1K Pot pressure has fallen to 7.5 mbar; tweak needle valve open additional 5 degrees. After 30 minutes it settles to 1.71 K / 20 mbar.
1558 Processing issues seem fixed. Set both phonon and charge prepulse bins to 500. Check phonon pulses:
DIB1pA Railed. Adjust Driver Offset.
DIB1pC Railed. Adjust Driver Offset.
DIB2pD Railed. Adjust Driver Offset.
DIB3pB Railed. Adjust Driver Offset.
DIB3pC Railed. Adjust Driver Offset.
DIB3pD Railed. Adjust Driver Offset.
1614 Commence LHe transfer (27.1%)
1653 Cease LHe transfer (90.4%).
1655 Check pulses:
DIB1pB Railed. Adjust Driver Offset. Now shows only pulseless traces. Adjust QET bias to 120uA. Pulses are back.
All other channels show good pulses.
1705 Fill LN2 cold trap full.
1722 Release pressure from Main Bath.
1745 Place Strong 88-Y source in Be disc sourceholder.
1747 Start 14 30min Y-Be series at -4/+4V with FG=5/2/5, OG=3, with 1000 initial randoms, 1000 ending randoms, 30s flash and 25 minute cooldown before series. Trigger on DIB1 phA with threshold of 20ADC bins and DIB1 qInner&qOuter each with threshold 210 ADC bins, with 150 us pileup cut window.