| | Setup: iZIP mode, DIB1/2/3 → cables → 3.2 MHz filters → DCRC55/33/50 |
| | Taking Cf data overnight. Will continue until ~1030 |
| 0905 | 1K Pot pressure has fallen to 3.2 mbar. Tweak needle valve open additional 5 degrees. |
| 0910 | Tower/MC(H35)/MC sensor/Cold Plate = 31.6/27.6/41/165 mK, Still = 1.37 K / 0.120 mbar, G1/G2 = 150/135 mbar, 1K Pot = 1.73 K / 20 mbar, Main bath = 70.7% |
| | Above numbers taken while still coming down from flash |
| 0920 | Fill LN2 cold trap Dewar. |
| 1200 | After dealing with computer issues, check all phonon pulses. Several of them (three on DIB1, one each on DIBs 2&3) were railed. Changing Driver Offset brings them back. Prepare for more noise tests today. |
| 1205 | Start test series with standard setup at 0/0 V with FG=5/2/5, OG=3, with 1000 initial randoms, 0 ending randoms, no flash. Trigger on DIB1 phA with threshold of 20ADC bins and DIB1 qInner&qOuter each with threshold 210 ADC bins, with 150 us pileup cut window. Stopping after randoms finish. |
| 1240 | Attach shielded pi-filter board to DCRC55. Switch all filter switches to OFF position. |
| 1307 | Check phonon pulses. All DIB1 channels railed, 1 DIB2, and 2 DIB3. Adjust driver offsets. |
| 1310 | Start test series with shielded pi-filter board with switches in OFF position at 0/0 V with FG=5/2/5, OG=3, with 1000 initial randoms, 0 ending randoms, no flash. Trigger on DIB1 phA with threshold of 20ADC bins and DIB1 qInner&qOuter each with threshold 210 ADC bins, with 150 us pileup cut window. Stopping after randoms finish. |
| 1320 | Unpower DCRC55, disconnect pi-filter board, and switch on filters for all pA related lines (QET A Return lines, QET A bias, SQUID Return, SQUID bias, Feedback Coil, Feedback Return): switches 1, 13, 14, 32, 17, 3, 20, 4, 21, 37. |
| 1345 | Reattach shielded pi-filter board and power DCRC55. Zap SQUIDs. |
| 1354 | Check phonon pulses. Only one channel on DIB2 railed. Adjust driver offset. |
| 1355 | Start test series with shielded pi-filter board with pA related switches in ON position, all others in OFF position at 0/0 V with FG=5/2/5, OG=3, with 1000 initial randoms, 0 ending randoms, no flash. Trigger on DIB1 phA with threshold of 20ADC bins and DIB1 qInner&qOuter each with threshold 210 ADC bins, with 150 us pileup cut window. Stopping after randoms finish. |
| 1505 | Disconnect pi-filter board. Reattach pi-filter connector and repower DCRC55. Zap SQUIDs. |
| 1513 | Check phonon pulses. One channel railed on DIB1, adjust Driver Offset. |
| 1516 | Start 15min test background series with standard setup at -4/+4 V with FG=5/2/5, OG=3, with 1000 initial randoms, 0 ending randoms, no flash. Trigger on DIB1 phA with threshold of 20ADC bins and DIB1 qInner&qOuter each with threshold 210 ADC bins, with 150 us pileup cut window. |
| 1535 | Stop run. Disable high-threshold on DIB1 Qi&Qo. |
| 1537 | Start 15min test background series with standard setup at -4/+4 V with FG=5/2/5, OG=3, with 1000 initial randoms, 0 ending randoms, no flash. Trigger on DIB1 phA with threshold of 20ADC bins, with 150 us pileup cut window. |
| 1554 | Stop run. |
| 1800 | Processing's all messed up, so just activate overnight. Place Cf source inside fridge shield near detector. Also flash detector since we never flashed after having it biased. |
| | Fill LN2 cold trap full. |