2016/01/19 (Tuesday)

Previous Day

Setup: iZIP mode, DIB1/2/3 → cables → 3.2 MHz filters → DCRC55/33/50
Taking Y-88 data overnight. Last flash at 06:32.
0935 Tower/MC(H35)/MC sensor/Cold Plate = 31.6/27.4/40/154 mK, Still = 1.15 K / 0.12 mbar, G1/G2 = 160/135 mbar, 1K Pot = 1.73 K / 19 mbar, Main bath = 68.1%
0940 Fill LN2 cold trap Dewar.
0945 Looking at pulses to see if they stayed overnight. Also eventually going to run a mini-IbIs to determine if QET biases are close to optimized.
All there, and all good. Slight ~200Hz noise.
0955 Perform “mini IbIs” calculations, observing phonon baseline voltages at varied QET biases: the standard bias and two large biases:
Vo (250 μA) Vo (300 μA) Ibias (μA) Vo (Ibias)
DIB 1 pA -0.212 -0.552 114.2 -0.696
pB 0.228 -0.094 127.1 -0.668
pC 0.274 -0.033 117.8 -0.714
pD -0.052 -0.3755 100.8 -0.645
DIB 2 pA -0.321 -0.5485 -112.6 -0.932
pB -0.1545 -0.3545 -112.6 -0.798
pC 0.108 -0.0745 -138.8 -0.860
pD -0.166 -0.4025 -97.7 -0.828
DIB 3 pA 0.763 0.514 110 -0.640
pB -0.166 -0.473 100.7 -0.788
pC -0.0725 -0.367 92.5 -0.552
pD 0.426 0.080 119.2 -0.631
note: for DIB2 (RevC) -250 and -300 μA were used

Resistances of TESs at the current QET bias settings are derived below following the method described on 2015 Sep 25:

Derived TES resistances (Ω)
DIB 1 DIB 2 DIB 3
pA 0.197 0.23 0.152
pB 0.19 0.246 0.173
pC 0.173 0.267 0.176
pD 0.169 0.186 0.157
1140 Set QET biases back to previous values. All the phonon channels show pulses. No obvious sign of 200Hz noise.
1148 Start 20 hours of 30min 88-Y series at +4/-4V with FG=5/2/5, OG=3, with 1000 initial randoms, 1000 ending randoms, 30s flash and 25 minute cooldown before series. Trigger on DIB1 phA with threshold of 30ADC bins.
1701 Fill LN2 cold trap full.