| | Setup: iZIP mode, DIB1/2/3 → cables → 3.2 MHz filters → DCRC55/33/50 |
| | Taking Y-88 data overnight. Last flash at 23:08. Why so early? Looks like DAQ stopped taking data too soon. |
| 0905 | Tower/MC(H35)/MC sensor/Cold Plate = 31.5/26.5/39/151 mK, Still = 1.10 K / 0.12 mbar, G1/G2 = 155/138 mbar, 1K Pot = 1.73 K / 19 mbar, Main bath = 32.9% |
| 0915 | Fill LN2 cold trap Dewar. |
| 0925 | Noticed that in the data beginning around noon yesterday the DIB2 phonon channels had very low noise amplitudes. In the ODB it appears that “PGAGain” for DIB2 phonons is set to 1; it should be set to 3. Although I'm not sure if the value to be entered is 3 or 8… register value 3 means 3 on RevC but it means 8 on RevC.1. Does MIDAS translate from input 8 down to register 3? The RevC.1 boards are set to 8 on the ODB. |
| 0943 | Start 30min 88-Y series at +4/-4V with FG=5/2/5, OG=3, with 1000 initial randoms, 1000 ending randoms, 30s flash and 25 minute cooldown before series. Trigger on DIB1 phA with threshold of 30ADC bins. |
| 1041 | Start 30min 88-Y series at +4/-4V with FG=5/2/5, OG=3, with 1000 initial randoms, 1000 ending randoms, 30s flash and 25 minute cooldown before series. Trigger on DIB1 phA with threshold of 30ADC bins. |
| 1205 | Pausing for some MIDAS test activity, so place Cf source in shield for activation while we're not taking data. |
| 1440 | Commence LHe transfer (23.3%). |
| 1526 | Complete LHe transfer (89.7%). |
| 1640 | After the MIDAS tests, check all the pulses. Looking good. |
| 1645 | Place Cf source back in safe. 88-Y source only one left inside shield. |
| 1657 | Start 17 hours of 30min 88-Y series at -4/+4V with FG=5/2/5, OG=3, with 1000 initial randoms, 1000 ending randoms, 30s flash and 25 minute cooldown before series. Trigger on DIB1 phA with threshold of 30ADC bins. |
| 1700 | Fill LN2 cold trap full. |