| | Setup: iZIP mode, DIB1/2/3 → cables → 3.2 MHz filters → DCRC55/33/50 |
| | Taking Cf data overnight. Last flash at 05:20. |
| 0850 | Tower/MC(H35)/MC sensor/Cold Plate = 33.6/27.9/41.4/162 mK, Still = 1.27 K / 0.12 mbar, G1/G2 = 125/115 mbar, 1K Pot = 1.73 K / 19 mbar, Main bath = 43.2% |
| 1004 | Start 30min Cf series at -4/+4V with FG=5/2/5, OG=3, with 1000 initial randoms, 1000 ending randoms, 30s flash and 25 minute cooldown before series. Trigger on DIB1 phA with threshold of 30ADC bins. |
| 1010 | Fill LN2 cold trap full. |
| 1102 | Start 30min Cf series at -4/+4V with FG=5/2/5, OG=3, with 1000 initial randoms, 1000 ending randoms, 30s flash and 25 minute cooldown before series. Trigger on DIB1 phA with threshold of 30ADC bins. |
| 1209 | Start 30min Cf series at -4/+4V with FG=5/2/5, OG=3, with 1000 initial randoms, 1000 ending randoms, 30s flash and 25 minute cooldown before series. Trigger on DIB1 phA with threshold of 30ADC bins. |
| 1307 | Start 2 cycles of 30min Cf series at -4/+4V with FG=5/2/5, OG=3, with 1000 initial randoms, 1000 ending randoms, 30s flash and 25 minute cooldown before series. Trigger on DIB1 phA with threshold of 30ADC bins. |
| 1525 | Commence LHe transfer (32.6%). |
| 1625 | Cease LHe transfer (92.7%). |
| | Fill LN2 cold trap full. |
| | Check all the channels before the overnight run. DIB2 pB was railed, so adjust Driver Offset. All the other channels were alive. Temporarily change trace length to 40ms, but no trace of the low frequency sinusoidal noise. Change the trace length back to 3.2768ms. |
| 1644 | Start 16 cycles of 30min Cf series at -4/+4V with FG=5/2/5, OG=3, with 1000 initial randoms, 1000 ending randoms, 30s flash and 25 minute cooldown before series. Trigger on DIB1 phA with threshold of 30ADC bins. |