start › cdms › k100 › run_diary › 20160106

2016/01/06 (Wednesday)

Previous Day

Setup: iZIP mode, DIB1/2/3 → cables → 3.2 MHz filters → DCRC55/33/50
Taking Cf data overnight. Last flash at 07:35.
0850 Tower/MC(H35)/MC sensor/Cold Plate = 33.1/27.4/41.4/166 mK, Still = 1.34 K / 0.12 mbar, G1/G2 = 116/112 mbar, 1K Pot = 1.71 K / 19 mbar, Main bath = 63.0%
1000 Fill LN2 cold trap full.
1015 Place Cf source back in safe.
1033 Start 2 cycles of 30min background series at -4/+4V with FG=5/2/5, OG=3, with 1000 initial randoms, 1000 ending randoms, 30s flash and 25 minute cooldown before series. Trigger on DIB1 phA with threshold of 30ADC bins.
1640 Commence LHe transfer (47.1%).
1720 Cease LHe transver (94.4%). Ich bin jetzt Deutsch.
1725 Fill LN2 cold trap full.
1745 Place Y-88 source inside shield at “optimal height” (~6“ down) for MISS events.
1815 MIDAS tinkering done. Restart frontends. All channels are in good health. No sign of fever.
Start 17 cycles of 30min Y-88 series at -4/+4V with FG=5/2/5, OG=3, with 1000 initial randoms, 1000 ending randoms, 30s flash and 25 minute cooldown before series. Trigger on DIB1 phA with threshold of 30ADC bins.