| | Setup: iZIP mode, DIB1/2/3 → cables → 3.2 MHz filters → DCRC55/33/50 |
| | Cf source near detector for overnight activation. |
| 0900 | Tower/MC(H35)/MC sensor/Cold Plate = 34.5/28.642.8/171 mK, Still = 1.36 K / 0.12 mbar, G1/G2 = 110/90 mbar, 1K Pot = 1.72 K / 20 mbar, Main bath = 20.6% |
| 0905 | Fill LN2 cold trap Dewar. |
| 0908 | Start LHe transfer at 17.3% |
| 1030 | Finish LHe transfer at 71.4% |
| 1200 | Still cannot get any channels (including charge) to unrail on DIB2. Cannot connect w/ Labview because of network issues. Cycle power to DCRCs. |
| 1215 | After power cycling, DIB2 shows pulses and charge channels are fine! Adjust some Driver Offsets so that the few railed channels unrail, but other than that, everything's looking good. |
| 1440 | Edit ODB so 1000 randoms are taken before and after run (as well as during run at a 1Hz rate). Enable writing of data. We are ready to run. Move Cf source up to top of dewar shield. |
| 1445 | Start 30min Cf series at -4/+4V with FG=5/2/5, OG=3, with 1000 initial randoms, 1000 ending randoms, 30s flash and 25 minute cooldown before series. Trigger on DIB1 phA with threshold of 30ADC bins. |
| 1549 | Start 30min Cf series at +4/-4V with FG=5/2/5, OG=3, with 1000 initial randoms, 1000 ending randoms, 30s flash and 25 minute cooldown before series. Trigger on DIB1 phA with threshold of 30ADC bins. |
| 1700 | Start 13 cycles of 30min Cf series at +4/-4V with FG=5/2/5, OG=3, with 1000 initial randoms, 1000 ending randoms, 30s flash and 25 minute cooldown before each series. Trigger on DIB1 phA with threshold of 30ADC bins. |
| 1705 | Fill LN2 cold trap full. |