2015/12/03 (Thursday)

Previous Day

Status: circulating overnight, base temperature, nothing plugged in to the DIBs
0820 Tower/MC(H35)/MC sensor/Cold Plate = 41.3/28.4/42.4/155 mK, Still = 1.02 K / 0.095 mbar, G1/G2 = 150/140 mbar, 1K Pot = 1.59 K / 5.5 mbar, Main bath = 36.5%
Tower temperature is still decreasing very slowly. Base doesn't seem quite as cold as the last run, MC or Tower. 1K Pot is running low. G1/G2 have been decreasing overnight from about 200 mbar.
0830 Open 1KPot needle valve about 5 degrees wider. 1KPot → 1.70 K / 17 mbar
0840 Fill LN2 cold trap Dewar
Tower temperature is decreasing a bit, maybe in response to 1KPot adjustment?
0854 Try heating still by 5 mW. Still P increases to 0.18 mbar, G1/G2 increase to 240/220 mbar. Tower temp spikes then falls, settling near its previous value of 40.7 mK.
0925 Turn off still heater
0930 Begin assembling shield
1015 1KPot pressure had fallen down to 8 mbar - open needle valve 5 degrees further.
1200 Tower/MC(H35) = 40.7/28.0 mK, Still = 0.10 mbar, G1/G2 = 190/170 mbar, 1K Pot = 1.68 K / 13 mbar, Main bath = 29.5%
1330 Finish assembling shield
1425 Plug in DCRCs same as last run: iZIP mode, DIB1/2/3 → cables → 3.2 MHz filters → DCRC55/33/50. Establish communication using telnet on jiajin, e.g. telnet 131.225.52.95 5000
1435 Start LHe transfer at 21.5%
1450 Start the three DAQ programs on the MIDAS webpage
1500 Try zapping with MIDAS. Doesn't look like it worked - after 5 minutes no change in Tower temperature or Still pressure.
1540 End LHe transfer (93.1%).
1635 Fill LN2 cold trap full.
1817 After some tests on MIDAS, set to flash overnight.