| | Setup: iZIP mode, DIB1/2/3 → cables → 3.2 MHz filters → DCRC55/33/50 |
| | overnight flashing detector. Tower temperature 150-170 mK. Temperature reading somewhat oscillatory. |
| 0910 | Tower/MC(H35)/MC sensor/Cold Plate = 160/52.3/69.3/163 mK, Still = 1.15 K / 0.14 mbar, G1/G2 = 195/175 mbar, 1K Pot = 1.80 K / 40 mbar, Main bath = 71.0% |
| | other thermometers seem to read more stably |
| 0915 | Fill LN2 cold trap Dewar |
| 0920 | Anthony stopped the bake. The bake was being done through MIDAS and should have stopped automatically earlier this morning, but for some reason it didn't. |
| 1135 | Tower/MC(H35)/MC sensor/Cold Plate = 39.2/28.2/42.4/160 mK, Still = 1.15 K / 0.13 mbar, G1/G2 = 185/165 mbar, 1K Pot = 1.80 K / 40 mbar, Main bath = 67.1% |
| 1500 | Tweaked 1KPot needle valve open a little more |
| 1605 | Tower/MC(H35)/MC sensor/Cold Plate = 38.7/29.2/43.7/168 mK, Still = 1.25 K / 0.13 mbar, G1/G2 = 175/155 mbar, 1K Pot = 1.77 K / 30 mbar, Main bath = 60.4% |
| 1820 | Put Cf source in fridge shield. Fill LN2 cold trap full. |