2015/11/30 (Monday)

Previous Day

0850 Status: pumping on IVC with turbo pump. MB in pit. Cf source in lead&poly castle activating detector since 2015/11/25.
0855 Put Cf source back in safe.
0900 IVC pressure pumping / with pump valved out = 5.4 / 340 e-6 mbar
1035 Place LN2 cold trap 1 back in bath and fill.
1045 Vacuum inside dewar and raise dewar.
1055 G1/G2/P1 = 121.8/560.5/2.6 mbar with all valves closed and pump off. Open V9 and turn on 3He pump, G2 rises to 777.8 mbar.
1100 Open V5 and V3 to pump P1 and G1 volumes to dump.
1135 G1: 0.6mbar, G2: 787.3mbar, P1: 15ubar.
1145 Commence cryocleaning mixture through Cold Trap 1. V1 open to 12.9%
1200 Adjust V1 to 13.1%
1300 Finish attaching Main Bath, all screws tightened
1310 Turn on He4 Rotary Pump and pump out Dewar.
1350 Open leak checker to IVC, get background leak rate of ~3-4 pbarL/s. Commence filling the main bath full of He. Leak rate increases slowly over next few minutes until main bath is full and leak rate is at ~10-13 pbarL/s. No sign of leak.
1356 Pump out 1K Pot.
1400 Add 4.6 mbar of He exchange gas to IVC.
1405 Commence LN2 transfer.
1425 Cernox MC thermometer = 295 K (down about 2 K)
1440 MC = 284 K, IVC = 4.0 mbar
1500 MC = 259 K, IVC = 3.4 mbar
1555 LN2 storage Dewar runs empty. Level resistor reads only 1.030 Ω when lowered into bath (0.985 Ω at room temperature). Need to refill storage Dewar.
1600 MC = 174 K, IVC = 2.2 mbar
1605 Cease cryocleaning mixture. Pump mix from Trap 1 to dump through valves 12a, 3, 6(100%), 9.
1615 Trap 1 down to G1/P1 = 0.6/0.085 mbar. Close V3 to isolate trap (monitoring its pressure on G1) and remove trap from LN2 to warm up.
1625 G2 (dump) reads 795.4 mbar. Close V6 and V9 and turn off 3He pump
1645 MC = 152 K, IVC = 2.2 mbar. Sorb hasn't changed, still reading 150 K
1650 G1 (Trap 1) has risen to 38.7 mbar, no longer frosty but still warming