2015/11/12 (Thursday)
| Current setup: HV Mode, DIB1/2/3 → cables → 3.2 MHz filters → HV Box → DCRC50/55. | |
| Overnight: idle, tower temperature stable. G2 falling slowly from 140 mbar to 130 mbar. | |
| 0820 | Tower/MC(H35)/MC sensor/Cold Plate = 30.5/24.3/38.3/139 mK, Still = 1.05 K / 0.11 mbar, G1/G2 = 120/130 mbar, 1K Pot = 1.68 K / 13 mbar, Main bath = 26.3% |
| Still, Cold Plate significantly cooler in HV setup, base somewhat cooler | |
| 0828 | Begin LHe transfer at 24.6% |
| 0830 | Fill LN2 cold trap Dewar. |
| 0927 | Finish LHe transfer at 90.4% |
| 0955 | Repeat some of yesterday's MC-heating measurements to see if we're in about the same place: |
| MC power [uW] | MC temp [mK] | Tower temp [mK] | CP temp [mK] | Still temp [K] | Still pressure [mbar] |
|---|---|---|---|---|---|
| 0 | 24.2 | 30.5 | 142 | 1.13 | 0.11 |
| 5 | 38.2 | 39.8 | 147 | 1.17 | 0.11 |
| 0 | 24.3 | 30.8 | 144 | 1.16 | 0.11 |
| interestingly the 0-power base is about the same but 5 uW doesn't heat it as much as yesterday | |||||
| CP and Still appear warmer but this is apparently a shift in the Fridge Out 1 temperatures… they shift back later at a random point | |||||
| 1135 | Apply HV voltage bias and monitor temperatures: |
|---|
| HV [V] | HV [nA] | MC temp [mK] | Tower temp [mK] | CP temp [mK] | Still temp [K] | Still pressure [mbar] |
|---|---|---|---|---|---|---|
| 0 | 0 | 24.3 | 30.8 | 144 | 1.16 | 0.11 |
| -10 | -9.6 | 24.2 | 31.3 | 144 | 1.15 | 0.11 |
| -10 | -9.2 | 24.2 | 31.3 | 144 | 1.15 | 0.11 |
| -15 | -19.6 | 24.5 | 32.6 | 137 | 1.04 | 0.11 |
| -15 | -19.5 | 24.6 | 32.7 | 137 | 1.04 | 0.11 |
| -20 | -30.1 | 24.7 | 33.9 | 136 | 1.04 | 0.11 |
| -20 | -30 | 24.8 | 33.9 | 136 | 1.04 | 0.11 |
| -30 | -51.9 | 25.3 | 37.2 | 136 | 1.04 | 0.11 |
| 0 | 0 | 24.3 | 30.8 | 136 | 1.04 | 0.11 |
| 10 | 11.5 | 24.2 | 31.2 | 136 | 1.04 | 0.1 |
| 10 | 10.5 | 24.2 | 31.2 | 136 | 1.04 | 0.11 |
| 20 | 32 | 24.5 | 33.4 | 137 | 1.04 | 0.11 |
| 30 | 52.8 | 24.9 | 36.2 | 136 | 1.04 | 0.11 |
| -40 | -74 | 25.3 | 39.5 | |||
| -50 | -96.1 | 26.1 | 43.6 | |||
| -75 | -148.9 | 27 | 51.5 | 137 | 1.04 | 0.1 |
| 50 | 97.5 | 26.2 | 42.6 | |||
| 75 | 152.3 | 27 | 50.3 | 138 | 1.07 | 0.11 |
| 0 | 0 | 24.9 | 31 | 137 | 1.04 | 0.11 |
| HVs of 30V or smaller are using the 0-40V supply; greater biases use the 0-1000V supply | ||||||
| 1425 | Return to iZIP mode, DIB1/2/3 → cables → 3.2 MHz filters → DCRC55/33/50. |
|---|---|
| 1500 | Try to set up everything from a Linux machine with MIDAS. Notes below: |
- Initiating communications with the board (i.e. the various Putty actions like wr 15 0, wr 15 20, sock, rd 30): on a Linux machine can ssh to jiajin then TELNET [DCRC address] [port# (use 5000)]. Then it's just like Putty.
- checked on the MIDAS browser, ODB→Equipment etc to see which ports it uses (5001 and 5002)
- on the MIDAS browser start the programs as in the instructions.
- Anthony knew this would apply charge biases, so went to ODB→Equipment→Towers→Charge to set the biases to zero
- from the browser go to DCRC_UI→tools→zap to zap. This worked well, one-touch, two flashes (could see them in still pressure)
- As per instructions, set up VNC session to jiajin and start the pulse viewing. Sadly this has some bugs currently, so stop there.
| 1625 | Set all phonon trigger thresholds to 65 ADC bins. |
|---|---|
| 1630 | Place Y-88 source inside shield. |
| 1735 | After taking back control from Midas, many phonon channels railed. Unrailing takes changing Driver Offsets. Many of the driver offsets that were set weren't consistent with what we had entered into MIDAS, so perhaps there's some discrepancy there. On DIB2, phB, C, and D are all showing upside down and dirty pulses. Similar to what we had seen on HV2pB earlier today. We haven't flashed since applying HV to the detectors, so maybe that's got something to do with it. Otherwise, we might be able to wait it out. |
| 1736 | Start 14 cycles of 30min runs of Y data at +/-4V with 30s of flashing and 25min cooldown beforehand, FG=5/2/5, OG=3, 1000 initial randoms, 2000 events/file, 30s flash and 25min cooldown in between. With all trigger thresholds set to 65ADC bins. |
| 1745 | Fill LN2 cold trap full. |
| 1746 | Flashing did not seem to help. Unplug DCRC33 and then repower. Zap twice and reapply settings. |
| 1754 | Without even having to stop DAQDAQ, having reapplied DCRC33's settings, all DIB2 phonon channels are showing proper pulses. Tada. |