| | Current setup: iZIP mode, DIB1/2/3 → cables → 3.2 MHz filters → DCRC55/33/50. |
| | Overnight: taking Y-88 data. Tower temperature between flashes stable. Last series began at 8:07 |
| 0815 | Tower/MC(H35)/MC sensor/Cold Plate = 31.0/25.5/40.8/156 mK, Still = 1.31 K / 0.11 mbar, G1/G2 = 125/130 mbar, 1K Pot = 1.68 K / 14 mbar, Main bath = 68.1% |
| 1000 | Disconnect DCRCs to install HV Box. |
| | New current setup: HV Mode, DIB1/2/3 → cables → 3.2 MHz filters → HV Box → DCRC50/55. |
| 1037 | Fill LN2 cold trap full. |
| 1250 | Perform a MC heating run, record temperatures at various stages: |