2015/11/13 (Friday)

Previous Day

Current setup: iZIP mode, DIB1/2/3 → cables → 3.2 MHz filters → DCRC55/33/50.
Overnight: taking Y data, last flash at 7:36.
0825 Tower/MC(H35)/MC sensor/Cold Plate = 31.0/25.4/39.0/143 mK, Still = 1.09 K / 0.10 mbar, G1/G2 = 115/110 mbar, 1K Pot = 1.64 K / 8 mbar, Main bath = 55.6%
1K Pot pressure and temperature have been falling steadily over the last 7 hours
0830 Fill LN2 cold trap Dewar.
0920 Notice that DIB2pB is railed. Looking at PSDs from overnight data it seems that something went wrong in the first series and it was dead throughout. Since this is Side2pA it's an outer phonon rail, so maybe not 100% critical… but mostly this is bad data.
1030 It looks like DIB2pB may have been in the state where a relock of the SQUID would temporarily unrail the channel, giving it the appearance of a good channel, but after a flash or changing of settings, the channel would then rail again, which is no good for overnight data. The proper and more permanent fix for this issue is to change the driver offset until a new value can be found, which keeps the channel unrailed for a more permanent period of time.
1230 Place Y-88 source inside shield at optimal height.
1234 Start 2 cycles of 30min runs of Y data at +/-4V, FG=5/2/5, OG=3, 1000 initial randoms, 2000 events/file, 30s flash and 25min cooldown in between. With all trigger thresholds set to 65ADC bins.
1310 1K Pot pressure & temperature continue to creep down, to 1.62 K / 7.0 mbar. Open needle valve about 5-10 degrees wider. Peaks to 1.8 K / 55 mbar, then oscillates a bit, stabilizing at 1.72 K / 22 mbar
1425 Looked at the difference between Tower temperature and MC temperature during two successive detector flashes:
time after flash start [m:s] Tower temp [mK] MC temp [mK]
0:42 192 45.5
0:54 47.3
0:30 108 45.0
2:00 40.5 35.0
5:00 34.5 30.1
bold indicates peak value
1447 Start LHe transfer at 44.3%
1520 Cease LHe trasnfer (91.6%).
1523 Start 43 cycles of 30min runs of Y data at +/-4V, FG=5/2/5, OG=3, 1000 initial randoms, 2000 events/file, 30s flash and 25min cooldown in between. With all trigger thresholds set to 65ADC bins.
1702 Fill LN2 cold trap full during a cooldown.