| | Current setup: iZIP mode, DIB1/2/3 → cables → 3.2 MHz filters → DCRC55/33/50. |
| | Overnight: Taking Cf data. Last flash at 7:22. Tower temp stable between flashes. |
| 0845 | Tower/MC(H35)/MC sensor/Cold Plate = 31.2/26.8/41.6/160 mK, Still = 1.39 K / 0.11 mbar, G1/G2 = 125/120 mbar, 1K Pot = 1.68 K / 13 mbar, Main bath = 19.8% |
| | G2 has fallen below G1, so basically back to normal. |
| 0850 | Fill LN2 cold trap Dewar |
| 0854 | Start LHe transfer at 19.4% |
| 0855 | Remove Soudan Cf source from fridge, put back in safe. |
| 0952 | Finish LHe transfer at 93.6% |
| 1030 | Set thresholds to October 21 values |
| 1036 | Start 30min run of background data at -/+10V overshooting +Qbias (side 2) to 14V before starting data, FG=5/2/5, OG=3, 1000 initial randoms, 2000 events/file, 30s flash and 25min cooldown after. Accidentally overshot wrong side so aborted this series early. |
| 1107 | Flash for about 1 minute using control vi |
| 1141 | Start 30min run of background data at -/+10V overshooting +Qbias (side 2) to 14V before starting data, FG=5/2/5, OG=3, 1000 initial randoms, 2000 events/file, 30s flash and 25min cooldown after. |
| 1254 | Start 30min run of background data at +/-10V overshooting +Qbias (side 2) to 14V before starting data, FG=5/2/5, OG=3, 1000 initial randoms, 2000 events/file, 30s flash and 25min cooldown after. |
| 1505 | Place Cf source inside shield near detector for Cf runs. |
| 1505 | Start 2 cycles of 30min runs of Cf data at -/+4V, FG=5/2/5, OG=3, 1000 initial randoms, 2000 events/file, 30s flash and 25min cooldown after. Using Wednesday's thresholds. |
| 1505 | Fill LN2 cold trap. |
| 1510 | Start 30 cycles of 30min runs of Cf data (alternating between +/-2V and +/-4V), FG=5/2/5, OG=3, 1000 initial randoms, 2000 events/file, 30s flash and 25min cooldown after. Using Wednesday's thresholds. |