2015/10/29 (Thursday)

Previous Day

Current setup: iZIP mode, DIB1/2/3 → cables → 3.2 MHz filters → DCRC55/33/50.
Overnight: Taking Cf data. Last flash at 8:16. Tower temp stable between flashes.
0910 Tower/MC(H35)/MC sensor/Cold Plate = 30.9/25.2/41.0/158 mK, Still = 1.33 K / 0.11 mbar, G1/G2 = 120/130 mbar, 1K Pot = 1.68 K / 13 mbar, Main bath = 62.3%
G2 is still higher than G1, so hard to say whether switching LN traps yesterday did anything. However the difference between the gauges has decreased over the last several hours from about 15 mbar to about 7 mbar.
0915 Put Soudan Cf source back in safe.
0920 Partially fill LN2 cold trap Dewar - storage Dewar ran empty.
0954 Place Ba source at back of fridge at topmost part of shield.
1000 Set phonon thresholds from yesterday's Ba runs:
A B C D
DIB1 56 60 62 60
DIB2 44 38 40 48
DIB3 60 66 78 68
1002 Start 30min run of Ba data at -/+8V overshooting +Qbias (side 2) to 14V before starting data, FG=5/2/5, OG=3, 1000 initial randoms, 2000 events/file, 30s flash and 25min cooldown after.
1102 Start 30min run of Ba data at +/-8V overshooting +Qbias (side 1) to 14V before starting data, FG=5/2/5, OG=3, 1000 initial randoms, 2000 events/file, 30s flash and 25min cooldown after.
1218 Start 30min run of Ba data at -/+6V overshooting +Qbias (side 2) to 14V before starting data, FG=5/2/5, OG=3, 1000 initial randoms, 2000 events/file, 30s flash and 25min cooldown after.
1318 Start 30min run of Ba data at +/-6V overshooting +Qbias (side 1) to 14V before starting data, FG=5/2/5, OG=3, 1000 initial randoms, 2000 events/file, 30s flash and 25min cooldown after.
1429 Fill LN2 cold trap full.
1430 Start 30min run of Ba data at -/+10V overshooting +Qbias (side 2) to 14V before starting data, FG=5/2/5, OG=3, 1000 initial randoms, 2000 events/file, 30s flash and 25min cooldown after.
1550 Start 30min run of Ba data at +/-10V overshooting +Qbias (side 1) to 14V before starting data, FG=5/2/5, OG=3, 1000 initial randoms, 2000 events/file, 30s flash and 25min cooldown after.
1633 Put Ba source back in safe.
1635 Place Cf source inside shield near detector for overnight Cf data.
1637 Set phonon thresholds back to Wednesday's values:
A B C D
DIB1 36 40 42 40
DIB2 24 18 20 28
DIB3 40 46 58 48
1653 Start 15 cycles of 30min runs of Cf data in the following order (-/+4V, +/-2V, +/-2V, +/-4V, +/-4V)X3, FG=5/2/5, OG=3, 1000 initial randoms, 2000 events/file, 30s flash and 25min cooldown after.
1728 Fill LN2 cold trap full.