| | Current setup: iZIP mode, DIB1/2/3 → cables → 3.2 MHz filters → DCRC55/33/50. |
| | Overnight: Taking background data. Tower temperature stable between flashes. |
| | Below readings taking after a flash, cooled most of the way to base: |
| 0900 | Tower/MC(H35)/MC sensor/Cold Plate = 31.1/25.9/42/161 mK, Still = 1.41 K / 0.11 mbar, G1/G2 = 145/145 mbar, 1K Pot = 1.72 K / 20 mbar, Main bath = 74.1% |
| | G1 and G2 nearly equal - is there an impedance building up in one of the traps? |
| 0905 | Fill LN cold trap Dewar. |
| 1445 | With MIDAS tests complete, place Cf source inside shield at top of shield for Cf data runs. |
| | Set thresholds to yesterday's values. |
| 1450 | Start Cf data : 22 cycles of 15min series at -/+2V, FG=5/2/5, OG=3, with 1000 initial randoms, 2000 events/file, 30s flash and 25min cooldown in between. |
| | Rate looking good. |