| | Overnight status: took background data, but DIB1pB went flatline during or after the first series |
| 0955 | Fill LN2 cold trap Dewar. |
| 1000 | G1/G2 much lower since about midnight for some reason, around 140 mbar |
| 1004 | Commence LHe transfer (30.1%) |
| 1048 | Finnish LHe tranfer (90.4%) |
| 1115 | After some MIDAS tests, reset for data taking. Several channels had to be relocked, one needed its offset adjusted. |
| | DIB2pD had Front-End Gain set to 5, should be 2 (mistake in MIDAS settings?) |
| | had to do wr 33 2000 on DCRC33(DIB2) to see charge. Actually isn't this only a problem for DCRC31? Maybe it was heating the FET that brought back the channels, not sure |
| | 250 Hz noise fairly low amplitude and intermittent. 2015/10/02 phonon thresholds look OK. |
| 1148 | Start 28 cycles of 15min background data series at -/+2V, FG=5/2/5, OG=3, with 1000 initial randoms, 2000 events/file, 30s flash and 30min cooldown in between. |
| 1155 | So far DIB1pB looks OK |