| | Current setup: iZIP mode, DIB1/2/3 → cables → 3.2 MHz filters → → DCRC55/33/50. |
| | Overnight: taking Cf data last night, last flash at 22:35. Bath is rather low: 1 K Pot temp and pressure have been falling, starting at about 6:20. |
| 0825 | Fill LN cold trap Dewar |
| 0830 | Tower/MC(H35)/MC sensor/Cold Plate = 31.0/25.4/45.1/159 mK, Still = 1.34 K / 0.080 mbar, G1/G2 = 175 mbar, 1K Pot = 1.70 K / 18 mbar, Main bath = 6.6% |
| 0850 | Start LHe transfer (6.2%). |
| 1025 | Cease LHe transfer (93.4%). |
| 1600 | SQUID tuning MIDAS program (see Pulse Viewing with MIDAS for instructions) was successfully used to view pulses and apply phonon settings. |
| 1620 | Use MIDAS to change trigger thresholds to the values set on 2015/09/15. |
| 1630 | Use MIDAS to apply -/+4 V charge bias through ODB. Assume that Qinner and Qouter values are applied correctly by MIDAS and do not need to be reversed, as is the case for Labview. Charge pulses were then checked using the MIDAS SQUID tuning GUI, and appeared normal. |
| 1640 | Use FlashandFlash from the DCRC_UI page to flash the detectors for 0.1s On and 5s off for 30s total. |
| 1650 | Fill LN2 cold trap full. |
| 1700 | Set 'Data write' to yes in MIDAS ODB/Logger in preparation for data taking. Start data run with MIDAS. |