2015/09/16 (Wednesday)

previous day

Current setup: iZIP mode, DIB1/2/3 → cables → 3.2 MHz filters → → DCRC55/33/50.
Overnight: no source; taking data overnight, last flash at 23:25 last night. All phonon channels still look good.
0905 Fill LN cold trap Dewar
0910 Tower/MC(H35)/MC sensor/Cold Plate = 31.0/25.3/41.0/158 mK, Still = 1.27 K / 0.080 mbar, G1/G2 = 155 mbar, 1K Pot = 1.72 K / 22 mbar, Main bath = 21.0%
1000 All phonon channels still showing regular pulses. Put Cf source in shield near detector in preparation for Cf data. Choose Top of Shield for source placement.
1005 Set phonon thresholds to same values as yesterday.
Set DAQDAQ2 to take:
  • 15 minute series of -/+2V then -/+4V data
  • 30s flash and 30min cooldown in between series
  • 2 cycles of these series
  • Frontend Gain for DIB1/2/3 set to 5/2/5
  • Output Gain set to 3 for all DIBs
  • OR trigger set across all phonon channels
  • 1000 initial randoms, 2000 events/file
1010 Start data series with Cf source. After randoms, rate starts to bog down. Quickly pull source up ~0.5 inches to position just underneath top of shield. Rate now ~12Hz.
1103 During data set that started at ~1057 rate drops to ~1.5Hz. Stop DAQDAQ. Increase phonon thresholds by 10 ADC bins each to lower trigger rate. Test run shows improved trigger rate of ~14Hz.
1115 Aborted data series: Run DAQDAQ for 1 more cycle of 15 minute runs of -/+2V and -/+4V data at same settings.
DIB2 pB railed. Stop run. Relock, but it doesn't help. Set offset back to 0.30 like yesterday, which brings the channel back at ~-0.7V.
1210 Midas test done. Check pulses. Some railed → reload settings. See strong ~300Hz sinusoidal noise on all DIBs. ~100mVpp on DIB3 and ~20mVpp on other two DIBs.
1215 Noise is weaker now even as I went through and measured the Vpp on all the DIBs. Leave it for a while longer till it “resolves” itself.
1230 Noise much less now (no longer sinusoidal).
Take Cf data: Start DAQDAQ for 2 cycles at same settings.
No channels railed during randoms. After randoms, event rate ~12Hz.
1600 Pulses look good. No large noise features.
1625 Commence LHe transfer (7.2%).
1735 End LHe transfer (72.1%).
1740 Take Cf data: Set DAQDAQ overnight to take 4 cycles of the -/+2V & -/+4V data series with same settings as before.
1810 Fill LN2 cold trap full.

next day