start › cdms › k100 › run_diary › 20150714

2015/07/14 (Tuesday)

previous day

Current configuration: DIB1/2/3 → 100MHz filter/none/none → DCRC55/33/50
0853 MC/Tower/Coldplate = 24.4/30.7/138 mK, Still = 0 mW / 0.11 mbar / 1.05 K, G1/G2 = 200/195 mbar, 1K Pot = 1.68 K / 13 mbar, Main bath = 22.5%
0854 G2 has been falling over the last few hours (198→195 mbar). Blockage forming in fridge? Nothing dramatic.
0906 Fill LN2 cold trap Dewar
0943 Power all three DCRCs, zap, set up phonons to previous settings
0945 looking at noise:
low frequency noise
DIB1 1 kHz @ 8/0/6/8 mVpp, 250 Hz @ 0/5/0/0 mVpp
DIB2 ~2 kHz in pC, 250 Hz @ 10/6/10/8 mVpp
DIB3 ~2 kHz in all, 250 Hz @ 5/3/6/0 mVpp
1020 Bake detector about 4 minutes.
1107 Tower has cooled to 32 mK.
1110 Visual assessment of pulse quality (poor/mediocre/decent/good) and noise:
pulse quality low frequency noise
DIB1 P/D/P/P 1 kHz @ 0/0/0/6 mVpp, 250 Hz @ 4/8/5/6 mVpp
DIB2 P/M/P/D ~2 kHz in pA & pC, 250 Hz @ 4/6/3/11 mVpp
DIB3 M/M/P/P ~1 kHz sporadic, 250 Hz @ 4/7/6/0 mVpp
1156 Observations: grounding DIB1 phonons improves DIB2 pulses, particularly DIB2pA and pC. Grounding DIB3 phonons doesn't help DIB2 much.
1204 Change output gain and driver offsets of DIB3 pC&pD, pC/pD = 1/1 = -0.088/-0.046. Previously these output gains were 0 unlike the other six RevC.1 phonon channels.
1209 Set thresholds:
Current thresholds
pA pB pC pD
DIB1 20 20 20 25
DIB2 25 20 20 15
DIB3 15 15 20 25
1220 failed data run, MultiDAQ was set for 1 DIB, restart it.
1231 Apply -4/4 V but MultiDAQ reset them to 0/0 V, start bkg data
1252 Stop data, 0/0 V, flash 30 seconds.
1310 Tower is hot, 37 mK. 250 Hz noise is large, DIB2 pC/pD at 30/40 mVpp. After a few minutes fridge drops to 33.5 mK (still high) and pD noise falls to 20 mVpp. looks like a Bad State.
1320 Disconnect all boards. Temperature fluctuations continue for over 30 minutes.
1405 Commence LHe transfer (13.8%).
1415 Tower down to 30.5 mK, reconnect boards & zap. Reset phonons.
1435 End LHe transfer (81.4%, 11 L remaining, dewar 3107).
pulse quality low frequency noise
DIB1 P/M/P/P very little
DIB2 P/P/M/D ~1 kHz on pC @ 4 mVpp
DIB3 M/D/P/P very little
1510 Apply -4/+4 V, start bkg data run
1528 Stop data, 0/0 V, flash 30 seconds.
1550 Apply -4/+4 V, start bkg data run
1607 Stop data, 0/0 V, flash 30 seconds.
1647 look at noise again:
low frequency noise
DIB1 1 kHz pD @ 5 mVpp, 200 Hz pD @ 4 mVpp
DIB2 1 kHz in pC @ 5 mVpp
DIB3 200 Hz pD @ 6 mVpp
1654 Apply -4/+4 V, start bkg data run. Rate bogs down to ~2 Hz at some point.
1607 Stop data, 0/0 V
1713 Bake detector
1752 Fill Ln2 cold trap full.

next day