start › cdms › k100 › run_diary › 20150715

2015/07/15 (Wednesday)

previous day

Current configuration: DIB1/2/3 → 100MHz filter/none/none → DCRC55/33/50. Baking overnight, no source.
0853 MC/Tower/Coldplate = 123/323/178 mK, Still = 0 mW / 0.15 mbar / 1.11 K, G1/G2 = 250/235 mbar, 1K Pot = 1.70 K / 13 mbar, Main bath = 61.0%
0848 Stop flashing detector.
0856 Fill LN2 cold trap Dewar.
0927 Tower falls to 32 mK
0930 MC/Tower/Coldplate = 26.4/31.7/145 mK, Still = 0 mW / 0.10 mbar / 1.11 K, G1/G2 = 215/200 mbar, 1K Pot = 1.69 K / 13 mbar, Main bath = 59.5%
pulse quality low frequency noise
DIB1 M/D/P/P 1-2 kHz @ 5/3/5/9 mVpp
DIB2 P/P/M/D ~1 kHz @ 6/4/5/3 mVpp
DIB3 P/D/P/P ~1 kHz @ 3/4/6/10 mVpp
1020 Estimated true event rate by looking for pulses in random 60 ms traces. Counted 45 events in 19 traces (1.14 seconds) yields event rate of 39 Hz. Note that this implies that about 16% of events will have pileup.
Current thresholds
pA pB pC pD
DIB1 25 25 20 30
DIB2 25 25 20 20
DIB3 20 20 15 25
1030 Start a data run but forgot to set voltage bias. Stop run.
1033 Set -4/+4 V. Start bkg data run.
1053 Stop data. Set 0/0 V. Flash 30 seconds.
1145 Less noise now:
pulse quality low frequency noise
DIB1 M/D/P/P none seen
DIB2 P/P/P/D ~1 kHz @ 4/0/3/0 mVpp
DIB3 P/D/P/P ~2 kHz @ 0/2/0/5 mVpp
1148 Set -4/+4 V. Start bkg data but data rate bogs down during randoms. Stop run.
1152 Try taking data again but bogs down again. Both times at about 800 randoms.
1203 Set to 0/0 V, experiment with taking data… Looking at the Readout vi, I noticed I had left the Qo trigger on with very low threshold to look at random noise. Reset the triggering to only the four phonons on every board using the Readout vi. This fixed the data rate problem. Apparently the Qo trigger was set on the DCRCs, and apparently this affects the data rate even during the randoms.
1207 Flash 30 seconds
Current thresholds
pA pB pC pD
DIB1 25 25 20 30
DIB2 25 25 20 20
DIB3 20 20 20 25
1243 200 Hz noise on DIB3 pD, 4 mVpp and DIB2pB, 4 mVpp.
1307 Stop data. Set 0/0 V. Flash 30 seconds.
1345 Background rate is about 20 events / 0.6 s = 33 Hz.
1350 Put Cf source near top of shield. Rate increases to about 24 events / 0.3 s = 80 Hz.
1355 Ensure trace length and triggers are set properly in Readout vi.
1356 Set -4/+4 V. Start Cf data run.
1414 Stop data. Set 0/0 V. Flash 30 seconds.
1504 Set -4/+4 V. Start Cf data run.
1524 Stop data. Set 0/0 V. Flash 30 seconds.
1548 Set -4/+4 V. Attempt run but LabVIEW error occurred involving enqueing a trigger… stop data.
1558 Start Cf data run. No errors.
1617 Stop data. Set 0/0 V. Flash 30 seconds.
1638 Set -4/+4 V. Attempt run but data rate bogs down after randoms. Stop data. Close Firefox and increase DIB3pC threshold to 20.
1644 Start Cf data run. Rate fine.
1702 Stop data. Set 0/0 V. Flash 30 seconds.
1703 Remove Cf source from fridge.
1725 Fill LN2 cold trap Dewar.
1730 Start baking detector.