2015/07/06 (Monday)

previous day

overnight status: Cf source near detector, baking detector. DIB1/2/3 → no filter/100 MHz C filter/3.2 MHz pi filter → DCRC55/33/50
0840 Tower = 298 mK, Still = 2 mW / 0.13 mbar, G1/G2 = 260/250 mbar, 1K Pot = 1.69 K / 13 mbar, Main bath = 72.1%
0841 Turn off flashing
0845 Fill LN cold trap Dewar
0955 DCRCs running, QETs at standard settings. Set new thresholds:
Current phonon threshold settings
pA pB pC pD
DIB1 25 25 30 20
DIB2 30 20 30 25
DIB3 25 15 10 25
Problems with B1 losing lock, requiring relock. Change B1 driver offset from -0.094 V to -0.08 V
note Oscillation seen in phonon channels ~175 Hz, 10/10/6 mVpp on DIB1/2/3.
1014 Set -4/+4 V. Start data, DIB3pB shifted: adjust driver offset from -0.012 to -0.098 V, stop data
1018 Start bkg data
1033 DAQ is sluggish, 100% CPU - stop run
1041 Thresholds still look fine, not too low. Test run with 100 randoms, CPU seems fine.
1046 Set 0/0 V, flash 30 seconds.
1058 Relock DIB1pB, apply -4/+4 V, start bkg data. Data taking runs fine.
1115 Stop data
1130 Scroll pump has been very loud. Prepare to stop using scroll pump as pot pump and try using the actual TRIVAC pot pump
  1. open V9 to allow mix to expand into dump
  2. close 1K Pot needle valve, close 1K Pot speedivalve, close V4a, close valve at pump-end of bellows
  3. switch off scroll pump, attach bellows to TRIVAC pot pump, start pot pump
  4. reopen pot valves in reverse order
  5. close V9, open V14 to allow mix back into dump
1148 V14 closed, all mix back in circulation. Base stayed pretty low during the switchover until opening V14, rose to over 100 mK. Adjusting 1K Pot needle valve to approximate previous pressure and temperature.
1240 Tower back to base of about 31 mK. Somewhat oscillatory, +/- 0.2 mK. Try doubling still power to 4 mW.
1345 Set to 0/0 V, flash long enough to hit stable temperature around 290 mK (about 3 minutes)
1455 Set -4/+4 V. Start bkg data
1514 Stop data, set 0/0 V, flash 30 seconds
1527 Set -4/+4 V. Start bkg data
1544 Stop data, set 0/0 V, flash 30 seconds.
1601 DIB1pB lost lock, relock
1602 Set -4/+4 (?) V. Start bkg data
1621 Stop data, set 0/0 V, flash 30 seconds. Note: DIB1 qI&qO had accidentally been set to -7 V at some unknown time
1623 Temperature is somewhat oscillatory with TRIVAC pot pump, try 5 mW still heat
1644 Still = 0.19 mbar, Tower 31.5 mK with ~0.2 mK oscillations. Try still heater at 0 mW
1650 Set -4/+4 V. Start bkg data
1707 Stop data, set 0/0 V, flash 30 seconds.
1735 Set -4/+4 V. Start bkg data. End with 30 second flash.
1742 Fill LN2 cold trap full.
1817 Set -4/+4 V. Start bkg data. Relock DIB1pB. End with 30 second flash.
1858 Set -4/+4 V. Start bkg data. Relock DIB1pB. End with 30 second flash.
1944 Set -4/+4 V. Start bkg data. Relock DIB1pB. End with 30 second flash.
2000 Start overnight bake.
2005 Tower at 295 mK. Put Cf source on fridge for overnight activation.