| 1021 | Relock DIB1pB again. Baseline shift in DIB1pD, change driver offset from -0.024 to -0.094. |
| 1024 | Set -4/+4 V. Start bkg data. |
| 1042 | Stop data, set 0/0 V, flash 30 seconds. |
| 1102 | Set to -3.5/+3.5 V. Start bkg data. (relock DIB1pB again) |
| 1116 | Fill LN2 cold trap Dewar |
| 1120 | Stop data, set 0/0 V, flash 30 seconds. |
| 1145 | Set to -3/+3 V. Start bkg data. |
| 1202 | Stop data, set 0/0 V, flash 30 seconds. |
| 1228 | Set to -2.5/+2.5 V. Start bkg data. (Had to relock DIB2pB and pC) |
| 1248 | Stop data, set 0/0 V, flash 30 seconds. |
| 1319 | Set to -2/+2 V. Start bkg data. (Had to relock DIB1pB) |
| 1339 | Stop data, set 0/0 V, flash 30 seconds. |
| 1407 | Set to -1.5/+1.5 V. Start bkg data. (Had to relock DIB1pB and DIB2pC) |
| 1426 | Stop data, set 0/0 V, flash 30 seconds. |
| 1449 | Set to -1/+1 V. Start bkg data. Note DIB1pB railed some time during run. |
| 1507 | Stop data, set 0/0 V, flash 30 seconds. |
| 1527 | Set to -0.5/+0.5 V. Start bkg data. |
| 1544 | Stop data, set 0/0 V, flash 30 seconds. |
| 1606 | Leave at 0/0 V. Start bkg data. Note DIB1pB railed some time during run. |
| 1651 | Commence He transfer. He level at 23.3%. |
| 1719 | Finish He transfer. He level at 77.3% (11 L left in dewar 3107). |
| 1730 | Fill LN2 cold trap Dewar |
| 1740 | Start baking detector. |
| 1815 | Take out Cf-252 source and place in shielding near detector for overnight activation. |