| 1040 | Set charge biases to -/+ 4 and start data. |
| 1050 | Fill LN cold trap full. |
| 1058 | Stop data, zero charge, flash. |
| 1111 | Start Ba data series. |
| 1129 | Stop data run, zero Q bias, flash. |
| 1140 | Start Ba data series. |
| 1158 | Stop data run, zero Q bias, flash. |
| 1208 | Start Ba data series. |
| 1230 | Stop data, zero charge bias, flash. |
| 1241 | Start Ba data series. |
| 1300 | Stop data run, zero Q bias, flash.; change thresholds to 25 ADC bins. |
| 1320 | Allison enters clean room for roughly 5 minutes to look at source holders. Opens purge cabinet on the right. |
| 1322 | Had to manually change trigger threshold for DCRC 50 through PUTTY. |
| 1323 | Start Ba data run. |
| 1340 | Stop data run, zero Q bias. |
| 1353 | Flash. |
| 1407 | Start Ba data run. |
| 1540 | Stop data run, zero Q bias, flash. |
| 1545 | Remove Ba source. Commence taking background data. |
| 1605 | Start bkg data run. |
| 1622 | Stop data run, zero Q bias, flash. |
| 1642 | Start bkg data run. |
| 1700 | Stop data run, zero Q bias. Zap Squids on DIB3 in hopes of fixing D3 which has weird dips in the pulses. |
| 1711 | All channels on DIB3 look “good” now, but D came back at an increased baseline (i.e., +0.4 V instead of -0.8 V). Change D3 driver offset from 0.028 V to -0.156 V. Had to relock C1 and D1. Change C1's driver offset from -0.016 V to 0.032 V. |
| 1720 | Save Multi-DAQ configuration in preparation to test DAQDAQ. |
| 1731 | Begin data run w/ DAQDAQ2; 3 data series of -/+4 V (10 min. each); only C1 shows a shifted baseline at ~ -0.4 V. |
| 1802 | During second series, A2 is railed at +1.2 V. Stop data run. Change A2 driver offset from 0.086 V to 0.032 V to unrail it. |
| 1807 | Flash w/ all 3 DIBs. |